Hose Connection on the COLDWELL

Filling the Labconco Reservoir

The Shorter KF-25 stub 

ALWAYS

 connects to the 

VACUUM PUMP

The Taller KF-25 stub 

ALWAYS

 connects to the 

VACUUM OVEN 

Make sure your have 800 ml of a heat transfer fluid such as ethanol, glycol or a cryocool inside the 

thermo reservoir. This fluid is sold separately.   

1. 

Remove the stainless COLDWELL. Pour the 800ml of heat transfer fluid into the 

 reservoir that the COLDWELL sits in.

2. 

Set the COLDWELL back inside the Labconco reservoir.

3. 

Monitor this fluid level during extended periods of use.

Make sure your have 800 ml of a heat transfer fluid such as ethanol, glycol or a cryocool inside the 

thermo reservoir. This fluid is sold separately.   

Summary of Contents for MAXTRAP 105

Page 1: ...s MAXTRAP 50 105 Supplement Instruction Manual Rev 11 29 18 www cascadesciences com info cascadesciences com 503 847 9047 Supplemental Installation Operation Manual MAXTRAP 50 105 Refrigerated Cold Tr...

Page 2: ...protection impairment by disabling or interfering with unit safety features This can result in damage to the unit and injury death or negative effects Follow all national laws regulations and local o...

Page 3: ...llowing information ready when calling or emailing Technical Support The model number and the serial number 503 847 9047 info cascadesciences com Cascade Sciences 6725 NE Evergreen Pkwy 106 Hillsboro...

Page 4: ...e period otherwise Cascade Sciences will not be held responsible Take photos of any shipping damage Refuse delivery or make note on the bill of lading 1 If anything is missing please call us at 503 84...

Page 5: ...separately from the MAXTRAP It is strongly recommend that the vacuum pumps GAS BALLAST be in the open position to best tolerate ethanol and other gas loads In some cases the vacuum pump can be powered...

Page 6: ...2 on the right side of the unit for venti lation and air to draw thru the slotted right side of the Thermo unit Electrical 120V The Labconco refrigerated base unit requires a standard 110 120 volt NEM...

Page 7: ...OVEN Make sure your have 800 ml of a heat transfer fluid such as ethanol glycol or a cryocool inside the thermo reservoir This fluid is sold separately 1 Remove the stainless COLDWELL Pour the 800ml o...

Page 8: ...l as important notes in the user manual Symbol Definition Consult the user manual Consulter le manuel d utilisation AC Power Rep re le courant alternatif AC Power Rep re le courant alternatif Begin he...

Page 9: ...not have to be under vacuum to operate Allow at least 30 minutes for the MAXTRAP fluid to reach operating temperature From The Vacuum Oven Vacuum valve is closed Vent Valve is closed Product is loade...

Page 10: ...ing Precautions 1 Do not move during operation 2 SAFETY TIP At very cold temperatures solvents and even oxygen can re condense back into liquid This is potentially explosive Especially if the cold tra...

Page 11: ...s that contain solvents capable of harming paint coatings or stainless steel surfaces Do not use chlorine based bleaches or abrasives these will damage the COLDWELL liner Vacuum Pump Maintenance Vacuu...

Reviews: