18 |
P a g e
INSTALLATION
R
EQUIRED
A
MBIENT
C
ONDITIONS
This oven is built for use indoors, at room temperatures between
15
°
C and 40
°
C (59
°
F and 104
°
F)
, at no
greater than
80% Relative Humidity
(at 25
°
C / 77
°
F). Operating outside these conditions may adversely
affect the oven temperature performance.
When selecting a location to install the unit, consider all environmental conditions that can adversely
impact its temperature performance. These include:
•
Proximity to other ovens, autoclaves, and any device that produces significant radiant heat
•
Heating and cooling vents or other sources of fast-moving air currents
•
High-traffic areas
•
Direct sunlight
R
EQUIRED
C
LEARANCES
These clearances are required to provide airflows for ventilation and cooling.
6 inches (152 mm)
of clearance is required on the sides.
12 inches (305 mm)
of headspace clearance is required between the top of the unit and any
overhead partitions.
Do not place objects on top of the oven.
Vacuum, backfill, and access ports are located on the back of the oven. Leave sufficient clearance
for operators to safely access these ports.
Gas In Ports
6” (152 mm)
12” (305 mm)
12” (305 mm)
12” (305 mm)
6” (152 mm)
Door Swing
24.8” (630 mm)
Auxiliary Port
Leave a 130° arc for the door swing. This
allows large sample trays to be removed
safely from the chamber without damaging
the door seal or metal sealing surfaces.
Summary of Contents for CTVR522-H
Page 1: ...Vacuum Control Oven SVO 5 VC Installation and Operation Manual ...
Page 6: ...6 P a g e TABLE OF CONTENTS ...
Page 16: ...16 P a g e RECEIVING ...
Page 28: ...28 P a g e SYMBOLS ...
Page 34: ...34 P a g e CONTROLS ...