24 |
P a g e
INSTALLATION
C
ONNECT
G
AS
S
UPPLIES
1. Connect your compressed air
supply
to this ¼-inch (6.35 mm)
push fitting.
•
The oven requires
70 – 80
psi
of pressure delivered at
this port to operate its
pneumatic valves
.
•
Never exceed 80 psi.
•
Failure to connect a
compressed air supply
prevents the oven from
pumping down or backfilling
the oven chamber.
Optional
: Connect equipment
to the KF-25 auxiliary port.
Comes with a blank and clamp.
Inert Backfill Gas Supply
Optional
: Connect gas
nitrogen or another inert
gas supply to this
3/8”
(9.5 mm) push fitting
for
backfilling the oven
chamber.
Do not exceed 15 psi of
pressure delivered at the
port.