CONTROLS
Oven Shelving Temperature Tab - Homepage
PV (Process Value):
•
When atmosphere is present in the chamber, the Process Value shows
the gas temperature.
•
When the oven chamber is pumped down, the Process Value shows the
current chamber shelving temperature.
SP (Setpoint) button
: A target, constant temperature the oven heats to and
maintains when a profile is not running. Setting the setpoint to zero effectively
gives the oven an unheated resting state.
PWR:
The power bar indicates the percentage of maximum power the
oven controller is calling for to heat the oven chamber.
Vacuum Control System Tab - Homepage
PV
: The present chamber pressure level, down to 1 Torr.
SP (Setpoint) button
: The Vacuum Control setpoint button. The oven pumps
down to this target when the Pump Start and Vacuum Control functions are on.
PWR:
Indicates the percentage of controlled air pressure to be used in
opening the oven vacuum valve. This pressure level is not applied until the
Vacuum Control function is turned on.
This power level is not used by the
Full Vacuum function.
The current oven chamber shelving temperature, 48.1°C
Chamber Pumping Down
Not Pumping Down
Oven heating with 100% power (PWR)
A Constant Temperature Setpoint (SP) of 150.0°C
entered on the setpoint button.
~760 Torr – Room
atmosphere
pressure at or near
sea level