50 |
P a g e
OPERATION
D
ATA
P
ORTS
Front of Unit
Control Panel USB A
The USB port located on the front control panel accesses the Main and High Limit controller and can
be used for the following:
•
Exporting and importing heating profiles from the controller
•
Updating firmware
Back of the Unit
Ethernet
The ethernet port located on the back of the oven accesses the Main and High Limit controller. With
the correct configurator software, the port can be used for the following:
•
Downloading profiles and uploading profiles to the controller.
•
Programming heating profiles in a desktop or laptop environment.
•
Backing up and loading backed up oven configurations.
O
VEN
C
OOLDOWNS
The oven chamber is well insulated and requires a significant amount of time to cool down while
remaining sealed.
•
6 hours may be required to cool down from 150°C to 100°C.
•
16 hours may be needed for the oven to return to room temperature cooling from 150°C.
•
Backfilling the oven with N
2
does not significantly increase the rate of cooling.
Introducing free atmosphere into the oven chamber at temperatures above 100°C risks oxidizing
chamber surfaces.