– 42 –
IT-9000
ALL TEST MENU-3
[1] LED Flash (Camera type only)
No.
Operation
Display
Test Details
1
Select “[1].LED Flash” from
MENU-3 screen.
When the “ENT” key is pressed,
the LED lights up. When it is
pressed again, the LED goes
out.
Then, the manual judgment
screen appears.
Lights properly (OK):
Press the “1” key to return to
the menu screen.
Does not light properly
(Failure):
Press the “0” key to return to
the menu screen.
[3] Low Voltage Detection
<Preparation>
(1) Open the battery pack cover and remove the battery pack.
(2) Connect equipment as shown below.
(3) Set the voltage control of the DC power supply at a minimum position, and then turn ON its
power.
DC power supply unit
DC voltmeter
Diode
(IF=3.0 A Max.)
Turn the right lock switch for the battery pack cover
to the “LOCK” position.
No.
Operation
Display
Test Details
1
Adjust the DC power supply
output so that the DC
voltmeter reads 7.4 V, and
then turn ON IT-9000.
Select “[3].Low Voltage
Detection” from MENU-3
screen.
Adjust the DC power supply
output so that the DC
voltmeter reads 5.5 V, and
then press the “ENT” key.
Automatically judges if XINHIN0
is not generated and then
IT-9000 automatically shuts off .
Turn ON IT-9000 again.
If XINHIN0 is not generated
(OK):
The screen returns to the menu
screen.
If XINHIN0 is generated
(Failure):
An error indication is displayed.
Press the “ENT” key to return to
the menu screen.