SKEMA 6
-
OPERATING INSTRUCTION
103
GB
103
GB
Operation of command for dental chair movements ( 3 ).
These buttons move the dental chair as follows:
-Chair seat up
-Chair back up
-Chair seat down
-Chair back down
To stop movement, release the button.
NOTE:
all dental chair movements are blocked when an instrument
is being used or the AUTOSTERIL system is running.
Right key ( 4 ) operation.
• Key held down (at least 2 seconds) with the instrument removed:
Water Clean System: sends a jet of running water to the instruments
such as the Turbine, Micromotor and Scaler to fl ush the spray ducts.
Water is delivered by pressing key ( 4 ); water is shut off when the key
is released and air is automatically blown to eliminate any drops of liquid
present in the spray ducts.
• Briefl y press key with instrument extracted:
Spray water to instrument on or off.
ATTENTION!
A beep sounds to signal the operating status has been changed.
LED (8
) lights up to signal operation with water spray.
Left key ( 5 ) operation.
• Key held down (at least 2 seconds) with the instrument removed:
Chip-air function: sends an air jet to the turbine or micromotor.
Air is delivered by pressing the button. Air is no longer blown when the
button is released.
• Briefl y press key with instrument extracted:
Spray water to instrument on or off.
ATTENTION!
A beep sounds to signal the operating status has been changed.
LED (8
) lights up to signal operation with water spray.
Right lever operation ( 6 ).
NOTE: the lever functions only with the instruments in their rest
position.
For safety reasons, the selected function starts only after the switch has
been briefl y actuated and then released.
• Lever pushed down:
"Dental chair automatic return” program activated.
• Lever pulled up:
Dental chair program “2" start.
Left lever operation ( 7 ).
NOTE: the lever functions only with the instruments in their rest
position.
For safety reasons, the selected function starts only after the switch has
been briefl y actuated and then released.
• Lever pushed down:
"Rinse position” (PR) program activated
NOTE: when the switch is actuated the second time, the dental
chair reaches its work position.
• Lever pulled up:
Dental chair program “1" start.
Protection against liquid penetration.
The foot control is protected against liquid penetration.
Degree of protection: IPX1.
Cleaning
Clean the foot control with a suitable product approved against bacteria and fungi and effective against viruses (including HBV and HIV): see
paragraph 1.4.
NOTE: The product should be sprayed on soft disposable paper towel or sterile gauze.
ATTENTION!
Do not use acids or harsh products.
NOTE: If the foot control slips on the fl oor, dust the slip-proof rubber found under the base with a dry cloth.