5.1.1.3.3. Emptying the M.W.B. system tank
NOTE: this menu can be accessed only if the M.W.B. system is
present.
This function is used to empty the M.W.B. system’s water circuit (see pa-
ragraph 7.3.) if the dental unit is not going to be used for several days or
if you want to drain the water present in the system.
Perform the following steps from the DISINFECTING SYSTEM SETTINGS
menu:
• Press button MENU to scroll the icons, selecting the one for "Emptying
the M.W.B. system tank".
• Press multifunction button OK to enter the selected submenu.
• Put the cup [ e ] provided under the spout.
• Press multifunction button (
) to start the the drain cycle.
NOTE: the drain cycle will not start if the S.S.S. system is turned
on or a fault is present in the M.W.B. system.
• Press multifunction button STOP to stop the drain cycle.
NOTE: the system automatically resets itself.
• Once the drain cycle has been completed, it is possible to shut off the
dental unit or press multifunction button ESC to reset the system to
restart work.
5.1.1.4. Hydrogroup settings
Perform the following steps from the main menu:
• Press button MENU to scroll the icons, selecting the one for "HYDRO-
GROUP SETTINGS" .
• Press multifunction button OK to enter the selected submenu that contains
the following icons:
Water to bowl settings
Water to cup settings
Automatic bowl motion settings