SKEMA 6
-
OPERATING INSTRUCTION
97
GB
97
GB
5.1.1.4.3. Automatic bowl motion settings
NOTE: this menu can be accessed only if the powered bowl is
present.
This setting is used for all the operators.
Perform the following steps from the HYDROGROUP SETTINGS menu:
• Press button MENU to scroll the icons, selecting the one for "Automatic
bowl motion settings" .
• Press multifunction button OK to enter the selected submenu that contains
the following icons:
Bowl rotation controller with rinse position retrieval for the dental
chair
Bowl rotation controller with automatic dental chair program
retrieval
Bowl
fl ushing controller with reset position retrieval for dental
chair
• Press button MENU to go from one icon to another.
• Use buttons INCREASE and/or DECREASE to select/deselect a con-
troller.
• To confi rm the selections made, simply quit this submenu by pressing
multifunction button
5.1.2. Setting the dental chair’s “rinse” and “home” positions
This setting is specifi c for each operator.
Perform the following operations from the main screen:
• Bring the dental chair into the desired position with the manual movements
buttons.
NOTE: in the case of the “Rinse position” the position of the bowl
can be saved, if powered.
• Hold button “SAVE” for at least 2 seconds to activate memory mode.
The system beeps and the relevant icon ( A ) appears on the console display
to signal save mode has been activated.
NOTE: Hold down button “MEMORIA” "SAVE" for at least 2 se-
conds to quit without saving the changes made.
• Press buttons “Automatic return” or “Rinse position” to assign the position
to the button.
The icon ( B ) for the selected program will appear on the console display
to confi rm the data have been saved.
NOTE: The seat height cannot be changed when in the “Rinse
position”.
NOTE: The “Rinse position” button brings the backrest and seat
into the rinse position.
When button “Rinse position” is pressed again, the backrest and seat return
to the previous position.