background image

Data logger operating manual

4

EN

6. Operation

The  PC  software  SmartGraph  is  the  central  configuration  interface  for  your 

data  logger.  All  additional  configuration  and  visualisation  specifications  can 

only be set by software. 

Basic settings can be directly configured with one-button operation by using

the mode selection button on your data logger. 

You can restrict the one-button operation with the mode selection button from

your  software  if  necessary  (key  lock).  It  is  not  possible  to  operate  your  data

logger with the mode selection button in this case.

6.1.    Switching on and off

When current is being supplied, the data logger cannot completely switch off,

but can only be set to an operating mode with minimal energy consumption

(M1). In this mode, measured value detection, measured value display and data

recording are inactive. An overview of the four various operating modes is pro-

vided in the next chapter. 

6.2.    Basic settings and operating modes

Seven  basic  settings  can  be  configured  by  using  the  mode

 selection  button.  These  include  the  four  various  operating

modes, the network function, a global reset of settings and

the acoustic function.

Briefly  pressing  the  mode  selection  button  changes  to  the

 current setting level. 

Briefly pressing the mode selection button allows navigation

through the individual setting modes. 

Each chosen mode is shown for four seconds and 

can be selected. 

Within this time, the mode marker flashes in the bottom left

corner of the display (M1, M2, M3, M4, M51, M52, M53).

Pressing the mode selection button for 1 second 

confirms your selection. 

The data logger then changes to the selected mode.

If no selection is confirmed within the four second period, the

setting level is left and the display returns back to the original

mode without making changes.

6.2.1.  The four operating modes

Operating mode M1

Measured value recording is inactive. “OFF” is shown in the second measured

value row. The STOP symbol is shown. 

In this operating mode (data logger delivery state), the power consumption is

minimal because measured values cannot be requested or shown.

M53

M52

M51

M4

M3

M2

M1

Selection of the operating mode M1

Display of the operating mode M1

5. Preparation before starting

5.1.    Software

5.1.1.  Installation conditions

To  configure  your  data  logger  and  read  the  recorded  measured  values,  the
SmartGraph software must be installed on a PC with the following minimum
 requirements.

Supported operating systems:

Windows XP from Service Pack 3 (32 bit or 64 bit version)

Windows Vista (32 bit or 64 bit version)

Windows 7 (32 bit or 64 bit version)

Hardware requirements:

Processor speed: 1 GHz, minimum 

CD-ROM drive

USB or network connection RJ45

512 MB RAM, minimum

4 GB of free hard disk space, minimum

Adobe Acrobat Reader software

5.1.2.  Installation of the SmartGraph software

Insert  the  CD-ROM  into  your  PC  drive  and  install  the  software  by  following 
the instructions in the installation wizard. 

5.1.3   Preparing data logger configuration

Connect  the  data  logger  to  your  PC  via  the  USB  cable  provided  in  the  scope 
of  supply.  The  measuring  device  is  automatically  detected  by  the  operating
 system.

Alternatively,  you  can  configure  your  data  logger  via  a  LAN  connection  over
your local network if the network function is enabled. Further information about
the network function is provided in chapter 6.2.2.

Start  the  SmartGraph  software.  The  program  automatically  detects  the  con -
nected data logger and adds it to the list of available data loggers. The data
logger can now be configured via the software.

Further  detailed  information  about  using  the  software  is  provided  in  the
 software manual which you can open from the help function of the Smart-
Graph software.

Functions of the professional version

Information about upgrading your SmartGraph software to the professional ver-
sion (improving the device license) is provided in the software manual. 

5.2.    Note at initial startup

After starting the device for the first time, the message “SET TIME”

appears on the display. However, no settings need to be made directly

on the device. The time synchronises itself with the PC time automatically when
connected to the SmartGraph software for the first time.

Summary of Contents for TCO

Page 1: ...BA THI THIP TCO 05 EN v1 0 Modification rights reserved Turfschipper 114 2292 JB Wateringen Tel 31 0 174 272330 www catec nl info catec nl Turfschipper 114 2292 JB Wateringen Tel 31 0 174 272330 www...

Page 2: ...table recording or re quest intervals then saved and transmitted to a connected PC The measuring device may only be used for this intended use while complying with specified technical data Any other u...

Page 3: ...Display symbol for power supply over a network Display symbol for power supply over USB Display for battery level Display for mode marker Display for active measured value recording Display for inacti...

Page 4: ...y state the power consumption is minimal because measured values cannot be requested or shown M53 M52 M51 M4 M3 M2 M1 Selection of the operating mode M1 Display of the operating mode M1 5 Preparation...

Page 5: ...on is ac tivated then software configuration and data reading from the data logger can be carried out over the network A signal which is broadcast from the data logger over UDP allows the SmartGraph s...

Page 6: ...minimum measured value min maximum meas ured value max and average measured value mid In total there are 24 meas uring channels available for your THI data logger as shown in table 1 Model THIP The TH...

Page 7: ...ed in the software manual 8 Alarm function 8 1 Alarm configuration An alarm function can be configured in the measuring device administration for each of the measuring channels selected for display or...

Page 8: ...ed batteries in the household rubbish or throw them in the fire instead dispose of them according to the relevant legal requirements 9 2 Deleting measured data The detected measured data are permanent...

Page 9: ...r at 25 C 0 5 hPa Screen resolution 0 1 hPa CO2 concentration Operating principle NDIR Measuring range 0 5 000 ppm Accuracy 50 ppm 3 measured value at 20 C and 1 013 mbar Screen resolution 1 ppm Long...

Reviews: