6.2.3.  Factory settings (M52)

This function resets the device settings
to factory settings. 

Even when resetting the device to factory
settings or when there are no batteries
in the device, the measurement data re-
mains in the memory and is not deleted.

Information about deleting the measured
data is provided in chapter 9.2.

6.2.4.  Acoustic function (M53)

Activating  or  deactivating  the  acoustic  function  switches  the  data  logger's
acoustic signal either on or off. If the acoustic function is active, the function's
display symbol is shown on the display.

When the acoustic function is active, alarm results are indicated as a tone, pro-
vided  an  active  alarm  has  been  preset  for  one  or  more  of  the  three  display
measured values in the SmartGraph software.

If the acoustic function is not active, then none of the navigation steps which
 require the mode selection button on the data logger to be pressed are acknowl-
edged  by  a  tone.  The  same  applies  for  selecting  a  mode.  If  no  selection  is
made, and thus the setting level is exited, then a tone is also emitted.

Selection of the reset function M52

Selection of the acoustic function M53

Example display of the acoustic function M53

7. Measured values display and 

data recording

7.1.    Sensors, channel groups and 

measuring channels

Model THI

The THI data logger has two internal sensors to detect measured values from a
total of six channel groups. These channel groups (measured values) are: Air
temperature  in  °C,  air  temperature  in  °F,  dew  point  in  °C,  dew  point  in  °F,
relative humidity in % and absolute humidity in g/m³.

For  each  channel  group,  there  are  four  measuring  channels  for  recording:
Current measured value (cur), minimum measured value (min), maximum meas-
ured value (max) and average measured value (mid). In total, there are 24 meas-
uring channels available for your THI data logger, as shown in table 1.

Model THIP

The  THIP  data  logger  has  three  internal  sensors  to  detect  measured  values
from a total of eight channel groups. These channel groups (measured values)
are: Air temperature in °C, air temperature in °F, dew point in °C, dew point in
°F, relative humidity in %, absolute humidity in g/m³, relative air pressure in hPa
and absolute air pressure in hPa.

For  each  channel  group,  there  are  four  measuring  channels  for  recording:
Current measured value (cur), minimum measured value (min), maximum meas-
ured value (max) and average measured value (mid). In total, there are 32 meas-
uring channels available for your THIP data logger, as shown in table 1.

Table 1: Overview of sensors, channel groups (measured values) and measuring channels of the data logger

Sensor / 
measured value sensor

Channel group 

(measured value)

Unit

Available measuring channels for data recording 

(max. 20 channels available for saving) 

and for displaying* (max. 3 channels for display)

Displayable 

in the display-

measured value

row

all 
models

internal 
temperature sensor 

Temperature

[°C]

cur

min 

max

mid

1, 2, 3

Temperature

[°F]

cur

min 

max

mid

1, 2, 3

Dew point

[°C]

cur

min 

max

mid

1, 2, 3

Dew point

[°F]

cur

min 

max

mid

1, 2, 3

internal 
humidity sensor

Relative humidity

[%]

cur

min 

max

mid

1, 2, 3

Absolute humidity

[g/m³]

cur

min 

max

mid

1, 2

only 
THIP

internal 
air pressure sensor

Relative air pressure**

[hPa]

cur

min 

max

mid

1, 2, 3

Absolute air pressure

[hPa]

cur

min 

max

mid

1, 2, 3

only TCO

internal 
carbon dioxide sensor

CO2 concentration**

[ppm]

cur

min 

max

mid

1, 2, 3

*      When selecting a channel for display, the current measured value (cur) is always automatically shown.
**    For more specific determination of measured values, entering the site height via the SmartGraph software is necessary.

Summary of Contents for TCO

Page 1: ...BA THI THIP TCO 05 EN v1 0 Modification rights reserved Turfschipper 114 2292 JB Wateringen Tel 31 0 174 272330 www catec nl info catec nl Turfschipper 114 2292 JB Wateringen Tel 31 0 174 272330 www...

Page 2: ...table recording or re quest intervals then saved and transmitted to a connected PC The measuring device may only be used for this intended use while complying with specified technical data Any other u...

Page 3: ...Display symbol for power supply over a network Display symbol for power supply over USB Display for battery level Display for mode marker Display for active measured value recording Display for inacti...

Page 4: ...y state the power consumption is minimal because measured values cannot be requested or shown M53 M52 M51 M4 M3 M2 M1 Selection of the operating mode M1 Display of the operating mode M1 5 Preparation...

Page 5: ...on is ac tivated then software configuration and data reading from the data logger can be carried out over the network A signal which is broadcast from the data logger over UDP allows the SmartGraph s...

Page 6: ...minimum measured value min maximum meas ured value max and average measured value mid In total there are 24 meas uring channels available for your THI data logger as shown in table 1 Model THIP The TH...

Page 7: ...ed in the software manual 8 Alarm function 8 1 Alarm configuration An alarm function can be configured in the measuring device administration for each of the measuring channels selected for display or...

Page 8: ...ed batteries in the household rubbish or throw them in the fire instead dispose of them according to the relevant legal requirements 9 2 Deleting measured data The detected measured data are permanent...

Page 9: ...r at 25 C 0 5 hPa Screen resolution 0 1 hPa CO2 concentration Operating principle NDIR Measuring range 0 5 000 ppm Accuracy 50 ppm 3 measured value at 20 C and 1 013 mbar Screen resolution 1 ppm Long...

Reviews: