54
Table 6.
Light indication of error codes. Reasons for appearance and methods of solution.
Error code
Error name
Reason
Solution
No signal from the
Chair angle sensor
Cable of the chair angle sensor are
disconnected or damaged
Reconnect the sensor cable and tighten the nut
Fault of the chair angle sensor
Contact the
Manufacturer
Fault of the electronics box
No signal from the
Base angle sensor
Fault of the electronics box
Contact the
Manufacturer
Chair or Base
angle sensor is
incorrectly
installed
Cable of the chair angle sensor are
disconnected or damaged
Reconnect the sensor cable and tighten the nut
Error of positioning of the chair angle sensor
or electronics box
Fix the sensor in its standard place
Fault of the chair angle sensor
Contact the
Manufacturer
Fault of the electronics box
High inclination of
the wheelchair
The wheelchair is turned over
Get off the obstacle
Wrong electronics box positioning
Fix the electronics box in its standard place
Fault of the electronics box
Contact the
Manufacturer
Overload of track
platform drive
High load on track platform
drive
Make sure that there is no obstacle when unfolding the
platform. Make sure the weight of the user, including items,
is less than 115 kg
Fault of the track platform
drive
Contact the
Manufacturer
Fault of the electronics box