Before getting near the compressor, cut the power off
YES
NO
EXTRAORDINARY MAINTENANCE RECORD
Compressed air distribution system
Record to be filled in at every extraordinary maintenance operation.
The extraordinary maintenance must be entrusted to a trained engineer,
provided with original spare parts .
At the first check up: check the screws of the terminals in the control panel
YES
NO
At the first check up: verify the head screw with a dynamometric key 6N.m
YES
NO
Every three months: clean the filters on the head (in a clean air environment)*
YES
NO
Every three months: verify the charging time (from 6 to 8 bar = 45/55 sec.)
YES
NO
Every three months: check the system and seek for any leak
YES
NO
Every three months: sterilize the absolute filter by autoclave
YES
NO
Every three months: verify the sound and visual signals
YES
NO
Every six months: replace the filters on the head (in a clean air environment)*
YES
NO
Every year: replace the absolute filter on the column
YES
NO
At every check-up: verify the functioning of ventilators
YES
NO
At every check-up: verify the functioning of drying system
YES
NO
At every check-up: verify the temperature of the technical room
YES
NO
At every check-up: check the noise
YES
NO
At every check-up: check the capacity of the capacitor
YES
NO
* In dusty environments clean and replace the filters more often, according to
environmental conditions
Date of maintenance ..................................... Operator’s signature ...........................................