5.3.3 – Re-Reading Stored Data in LOG - Measure IF (Infrared)
To read again these data proceed as follows:
Keeping pressed the measurement trigger (Ref. Chap.3 – Pos.9), press repeatedly the key
MODE
until on the
display it appears the abbreviation
LOG
(Ref. Chap.4 – Pos. 14), at this point release the trigger, and operating
on the keys
UP
&
DOWN
, it is possible to scroll the several memory positions, visualizing them from 1 to 100
on the display in little on the high side (Ref.Chap.4 – Pos.8) , the stored temperature corresponding to the
memory cell is visualized on the display in little on the low side (Ref. Chap.4 – Pos.9)
Note: Not touching anymore any key for at least 7 seconds the instrument switch itself off
5.3.4 – Erasing of the LOG Storage
This function allows to erase quickly all the stored data in the LOG, and to proceed execute the following
procedure:
The erasing can be made only when the instrument is in
LOG
, this state is confirmed by the visualization of the
abbreviation on the (Ref. Chap. 4 – Pos. 14), if the display does not visualize this abbreviation, always in
MEASURE
, with pressed trigger, press repeatedly the key
MODE
(Ref.Chap.5 – Pos.4) until it appears the
abbreviation LOG.
Once assured to be in
LOG
, always keeping pressed the trigger, pressing repeatedly the key
DOWN
up to
arriving in visualizing the memory cell
0
, visualized on the display little in the high side (Ref.Chap. 4 – Pos.8),
without releasing the trigger, press shortly the red key Backlighting/Laser (Ref. Chap.5 – Pos.1), the
instrument will confirm the erasing with a double sound note, and the display will visualize the memory cell
1
,
ready to begin a new storage
5.3.5 – Activation –Deactivation USB Gate to download data
The instrument is supplied by a digital communication USB gate, through which it is possible to download the
stored data to visualize them on the PC.
Activation USB Gate: To activate it, it is necessary that the instrument is in the state
MAX
-
MIN
-
DIF
-
AVG
, with
one of these abbreviations visualized on the display (Ref. Chap.4 – Pos.7), then keep pressed for few seconds
the red key Backlighting-Laser (Ref. Chap.5 – Pos.1) until the abbreviation
USB
appears in the low right corner
of the display (Ref. Chap.4 – Pos.11).
Deactivation USB gate: To deactivate the function, it is necessary that the instrument is in the state
MAX
-
MIN
-
DIF
-
AVG
, with one of these abbreviations visualized on the display (Ref. Chap.4 – Pos.7), then keep pressed for
few seconds the red key Backlighting-Laser (Ref. Chap.5 – Pos.1) until the abbreviation
USB
disappears from
the low right corner of the display (Ref. Chap.4 – Pos.11).
Summary of Contents for IF324
Page 2: ...IF324_5A808_UK_M1 Copyright CEAM Group CEAM Control Equipment 2 ...
Page 4: ...IF324_5A808_UK_M1 Copyright CEAM Group CEAM Control Equipment 4 ...
Page 5: ...IF324_5A808_UK_M1 Copyright CEAM Group CEAM Control Equipment 5 ...
Page 21: ...IF324_5A808_UK_M1 Copyright CEAM Group CEAM Control Equipment 21 ...
Page 22: ...IF324_5A808_UK_M1 Copyright CEAM Group CEAM Control Equipment 22 ...
Page 23: ...IF324_5A808_UK_M1 Copyright CEAM Group CEAM Control Equipment 23 ...