background image

Gauges are calibrated at the factory using equipment traceable to 
NIST. There is no need to calibrate the gauge before putting it into 
service. Calibration intervals depend on your quality control pro-
gram requirements, although many customers calibrate annually.
Calibration should only be performed by qualified individuals using 
appropriate calibration standards and procedures. The calibration 
equipment should be at least four times more accurate than the 
gauge being calibrated. 
The calibration system must be able to generate and measure pres-
sure/vacuum over the full range of the gauge. 
A vacuum pump able to produce a vacuum of 100 microns (0.1 
torr or 100 millitorr) or lower is required for vacuum and absolute 
gauges. 
Warning: application of vacuum to non-vacuum models may result 
in irreparable damage to the sensor.
Allow the gauge to acclimate to ambient temperature for 20 
minutes. 
Remove the rear cover to gain access to the UP and DOWN buttons 
located near the lower right and left corners of the circuit board.
Install fresh batteries.

Entering Calibration Mode

With the gauge off, press and hold the DOWN button. Then press 
the front button. Release all buttons when the display indicates CAL.
The display begins by indicating the full-scale positive pressure 
rating of the gauge in the engineering units as configured by the 
factory, and then shows all display.
Before the gauge enters the Calibration Mode, the display initially 
indicates _ _ _ _ with the first underscore blinking, and with CALPC 
(calibration passcode) on the lower display. 
Note: The gauge will automatically revert to normal operation if 
no buttons are operated for approximately 15 seconds. To cancel 
and return to normal operation, press and release the front button 
without entering any passcode characters.

Enter the User-Modifiable Passcode

The factory default is 3510, but this is user changeable.
1.  Use the UP or DOWN buttons to set the left-most digit to 3.
2.  Press and release the front button to move to the next position. 

The 3 will remain, and the second position will be blinking.

3.  Use the UP or DOWN buttons to select 5.
4.  Press and release the front button to index to the next position. 

35 will remain, and the third position will be blinking.

5.  Use the UP or DOWN buttons to select 1.
6.  Press and release the front button to index to the next position. 

351 will remain, and the fourth position will be blinking.

7.  Use the UP or DOWN buttons to select 0.
8.  Press and release the front button to proceed with configuration 

procedures. 

If an incorrect passcode is entered, the gauge will return to the start 
of the passcode entry sequence.

Calibration Mode

The gauge enters and remains in the calibration mode until 
restarted manually or power is removed. Features not related to 
calibration are disabled and compound range models are set for the 
same engineering units for pressure and for vacuum.
The calibration may be performed in any of the available engineer-
ing units as well as percent (PCT). For greatest accuracy, use the UP 
and DOWN buttons to select engineering units for calibration with 
highest resolution (highest number of display counts). Press and 

Instructions

CTP1B

release the front button when the appropriate engineering units are 
displayed. Suggested units are listed below.
Sensor 

Suggested units for calibration

5 PSI 

5.000 PSI

15 PSI 

775.7 MMHG  (TORR)

30 PSI 

61.08 INHG

50 PSI 

50.00 PSI

60 PSI 

60.00 PSI

100 PSI 

7.031 KG/CM2

200 PSI 

407.2 INHG

300 PSI 

610.8 INHG

500 PSI 

500.0 PSI

1000 PSI 

70.31 KG/CM2

2000 PSI 

4072 INHG

3000 PSI 

6108 INHG

5000 PSI 

5000 PSI

Any 

100.00 PCT (percent)

The display will then indicate the currently applied pressure in the 
engineering units selected for calibration.

UP and DOWN Button Operation

Each time one of the calibration buttons is pressed and released 
quickly, a small change is made to the digitized pressure signal. It 
may take more than one of these small changes to result in a single 
digit change on the display. 
To make larger changes, press and hold the appropriate calibration 
button. After about one second, the display will begin to change 
continuously. Release the button to stop. Then make fine adjust-
ments by pressing and quickly releasing the calibration buttons as 
previously described. 

Gauge Reference Pressure Gauges

Apply zero pressure by venting the gauge port to atmosphere. The 
character display will alternate between ZERO and CAL. Adjust for 
a display indication of zero using the UP and the DOWN buttons.
Apply full-scale pressure. The character display will alternate 
bSPAN and CAL. Adjust for a display indication of full-
scale pressure using the UP and the DOWN buttons. 
Apply 50% full-scale pressure. The character display will alternate 
between  +MID and CAL. Adjust for a display indication equal to 
50% of full-scale pressure using the UP and the DOWN buttons. 

Gauge Reference Vacuum Gauges

Apply zero pressure by venting the gauge port to atmosphere. The 
character display will alternate between ZERO and CAL. Adjust for 
a display indication of zero using the UP and the DOWN buttons.
Apply full-scale vacuum. The character display will alternate 
bSPAN and CAL. Adjust for a display indication of full-
scale vacuum using the UP and the DOWN buttons. 
Apply 50% full-scale vacuum. The character display will alternate 
bMID and CAL. Adjust for a display indication equal to 
50% of full-scale vacuum using the UP and the DOWN buttons. 

Absolute Reference Gauges

Apply full vacuum to the gauge. The character display will alternate 
between ZERO and CAL. Press the UP and DOWN buttons to obtain 
a display indication of zero.
Apply full-scale pressure. The character display will alternate 
bSPAN and CAL. Press the UP and DOWN buttons to obtain 
a display indication equal to full-scale pressure.
Apply 50% of full-scale pressure. The lower display will alternate 
bMID and CAL. Press the UP and DOWN buttons to obtain 
an indication equal to 50% of full-scale pressure.

Compound and Bipolar Gauges

In addition to the steps described above for pressure gauges, apply 
full-scale vacuum. The character display will alternate between   
–SPAN and CAL. Adjust for a display indication of actual applied
vacuum using the UP and the DOWN buttons.
For bipolar and –30.15.00psig compound range models 
only, apply 50% full-scale vacuum. The character display will alter-
nate between –MID and CAL. Adjust for a display indication equal 
to 50% of full-scale vacuum using the UP and the DOWN buttons. 

Save Calibration

Once the adjustments are complete, press and hold the front but-
ton until the display indicates  – – – –  then release the button to 
store the calibration parameters in non-volatile memory and restart 
the gauge.
Verify the pressure indications at 0%, 25%, 50%, 75% and 100% 
of full scale.
Replace the rear cover taking care not to pinch the wires between 
the cover and the case.

Remove the rear cover to access the buttons located near the lower 
right and left corners of the circuit board.

View Or Change User Configuration Passcode

With the unit off, press and hold the UP button, then press the 
front button. 
Release all buttons when the display indicates CFG.

View Or Change User Calibration Passcode

With the unit off, press and hold the DOWN button, then press the 
front button.
Release all buttons when the display indicates CAL.

Enter Access Code 1220

Before the unit enters the view or change passcode mode, the 
display initially indicates  _ _ _ _  with the first underscore blinking, 
and with CFGPC or CALPC on the character display.
Note: The gauge will automatically revert to normal operation if no 
buttons are operated for approximately 15 seconds. 
To cancel and return to normal operation, press and release the 
front button without entering any passcode characters.
1.  Use the UP and DOWN buttons to set the left-most digit to 1.
2.  Press and release the front button to move to the next position. 

The 1 will remain, and the second position will be blinking.

3.  Use the UP and DOWN buttons to select 2.
4.  Press and release the front button to index to the next position. 

12 will remain, and the third position will be blinking.

5.  Use the UP and DOWN buttons to select 2.
6.  Press and release the front button to move to the next position. 

1 2 2 will remain, and the fourth position will be blinking.

7.  Use the UP and DOWN buttons to select 0.
8.  Press and release the front button to proceed. 
Note: If an incorrect access code was entered, the gauge will return 
to the start of the access code entry sequence.
Once the access code has been entered correctly, the display will 
indicate the existing user-defined passcode with CFGPC or CALPC 
on the character segments.
1. Operate the UP or DOWN button to select the first character of

the new passcode.

2. When the correct first character is being displayed, press and

release the front button to proceed to the next passcode char-
acter.

3.  Repeat above until the entire passcode is complete.
4. To exit, press and hold the front button. Release the button

when the display indicates – – – – to restart the gauge.

5. Replace the rear cover taking care not to pinch the power wires 

between the cover and the case.

User-Defined Passcode Configuration

Calibration Preparation

Calibration

Calibration—continued

www.mod-tronic.com | sales@mod-tronic.com | 1-800-794-5883

Reviews: