background image

Instructions CTP3B

Enter Configuration Passcode 

Enter the passcode. 3510 is the factory 
default, but it is user-modifiable.
Use the 

s

 or 

t

 buttons to set the left-

most digit to 3.

Press and release the Power button to 
index to the next position. 
The 3 will remain, and the second posi-
tion will be blinking.

Use the 

s

 or 

t

 buttons to select 5.

Press and release the Power  button 
to index to the next position. 3 5 will 
remain, and the third position will be 
blinking.
Use the 

s

 or 

t

 buttons to select 1.

Press and release the Power  button 
to index to the next position. 3 5 1 will 
remain, and the fourth position will be 
blinking.
Use the 

s

 or 

t

 buttons to select 0.

Press and release the Power button to 
proceed with configuration. 
Note: If an incorrect passcode is entered, 
the gauge will return to the start of the 
passcode entry sequence.

Press and release the Memory button 
until Max or Min is displayed.
Depending on your configuration the 
gauge may save both Max and Min, or 
either Max or Min.
The gauge may be left in the maximum 
or minimum display mode if desired. The 
reading will be continuously displayed, 
stored and updated.
The center power button may be pressed 
at any time to return to the normal dis-
play mode.

Min/Max Memory 

The Min/Max readings are captured at the rate of 3 times per 
second. If a brief pressure deviation occurs, it may not be cap-
tured. The readings are captured any time the gauge is on, but 
not while in the configuration or calibration mode. 
The gauge can be configured to capture and display:

 

O

Both maximum and minimum values (factory default)

 

O

The maximum value only

 

O

The minimum value only 

 

O

Saved Min/Max values when the gauge powers down

Min/Max Memory Retention

The upper display indicates 

clr

 and the 

lower display will indicate the current 
setting for min/max memory. Press: 
Power button: Keep setting

s

 button: Select AUTO and automatically 

clear the maximum and/or minimum 
readings when the gauge is shuts off. 

t

 button: Select MAN and retain the 

maximum and/or minimum readings in 
memory after the gauge is shut off. The 
readings can be cleared manually.
Press and release the power button to 
save your selection and move to the next 
parameter.

cecomp.com

1220 American Way Libertyville, IL 60048 

800-942-0315

CECOMP

 

Div. of

 

A

A

BSOLUTE

BSOLUTE

 P

 P

ROCESS

ROCESS

 I

 I

NSTRUMENTS

NSTRUMENTS

GAUGE CONFIGURATION 

Enter Configuration Mode 

The gauge uses a 4 digit passcode 
to enter the configuration mode. This 
is to prevent unauthorized changing of 
settings.
With the gauge off, press and hold the 

button.

Then press the center power button.
Release all buttons when the display 
indicates 

CFG

. The gauge firmware ver-

sion is also displayed.
The gauge then goes through the normal 
power up sequence.
The display prompts for entry of the 
configuration passcode (CFGPC), with the 
first underscore blinking. 

Note: The gauge will automatically revert 
to normal operation if no buttons are 
pressed for approximately 15 seconds. 
To cancel and return to normal operation, 
press and release the power button with-
out entering any passcode characters.

hold

Zero/Clear

Memory

G

F

C

firmware version

power up sequence

_

_

_

_

CFGPC

t

s

*

1

*

2

Zero/Clear

Memory

_

_

_

3

CFGPC

- or -

_

_

5

3

CFGPC

Zero/Clear

Memory

_

_

_

3

CFGPC

_

I

5

3

CFGPC

Zero/Clear

Memory

0

I

5

3

CFGPC

t

s

*

*

t

s

*

t

s

*

r

l

c

Auto

man

Zero/Clear

Memory

t

s

or

*

*

Engineering Unit Selection 

Note: For vacuum/pressure gauges or 
compound gauges the choice of CMPND 
(inHg/psi) or -/+EU (±Engineering Units) 
will appear. Select -/+EU to change to 
other engineering units. 
Engineering units are shown in the lower 
display.
Use the 

s

 and 

t

 buttons to navigate 

through the list of engineering units. 
Available engineering units depend on 
the sensor range.
When the desired units are displayed, 
press and release the Power button to 
save your selection and move to the next 
parameter.

Zero/Clear

Memory

PSIG

INHG

Zero/Clear

Memory

t

s

*

*

- or -

t

s

*

Auto Shutoff Time Selection 

The auto shutoff time is displayed on 
the upper display. The lower display 
will indicate AST M for minutes or AST 
H for hours. 
Use the 

s

 and 

t

 buttons to select 0 

(manual shutoff), 1, 2, 5, 10, 15, 20, or 
30 minutes, or 1, 2, 4, or 8 hours.
A setting of zero disables the auto shutoff 
timer. Use the Power button to shut the 
gauge off.
When the desired time is displayed press 
and release the Power button to save 
your selection and move to the next 
parameter.

Clear a Memory Location 

For many applications it may be desir-
able to bring the system up to normal 
pressure and then clear the minimum or 
maximum value.
Press and release the Memory button 
until the value to be cleared is displayed.

Press and hold the Zero/Clear button. 

Release the button when 

clr

 is dis-

played. The reading for the indicated 
memory location will be cleared.
With a gauge reference models if no 
pressure is applied, the value will return 
to zero.
If pressure is applied the new pressure 
reading will be stored in memory.
Absolute reference models will store the 
current atmospheric pressure reading if 
the gauge port is open to atmosphere.
Press and release the Power button to 
exit the memory mode and return to live 
pressure readings.

0

0

 MAX

Zero/Clear

Memory

t

s

*

i90.0

Zero/Clear

Memory

*

t

s

MIN  MAX

or

0

0

 PSIG

Zero/Clear

Memory

t

s

*

i92.0

MIN  MAX

or

r

l

c

 MAX

MAX  min

i00.0

psig

i90.0

i92.0

 Max

Zero/Clear

Memory

*

Zero/Clear

Memory

t

s

t

s

*

or

Continued >>

Backlight Time Selection 

The upper display will indicate 

bL

 to 

indicate display backlight setup.
Use the 

s

 and 

buttons to select

AUTO: backlight enabled for 1 minute.
ON: backlight on when the gauge is on.
OFF: display backlight is disabled.
When the desired setting is displayed, 
press and release the power button to 
save your selection.  
The gauge will restart and is ready to 
use with your new configuration.

AST M

AST H

Zero/Clear

Memory

Zero/Clear

Memory

t

s

*

*

- or -

t

s

*

I

5

Zero/Clear

Memory

AUTO

on

Zero/Clear

Memory

auto

t

s

*

*

- or -

t

s

off

*

L

p

L

p

Min/Max Setup 

The  lower display will indicate the cur-
rent MX/MN configuration. 
Press the power button to keep it, or use 
the 

s

 or 

t

 buttons to select the desired 

configuration.
MX/MN Capture both maximum and 

 

 

minimum readings

--/MN   Capture minimum readings only
MX/--  Capture maximum readings only
Press and release the power button to 
save your selection and move to the next 
parameter.

MX/--

--/MN

MX/--

MX/MN

Zero/Clear

Memory

Zero/Clear

Memory

t

s

*

*

- or -

t

s

*

Gauge Configuration—User or Factory 

Upon successful passcode entry, the lower display will indicate 
USER.
To reset the configuration to factory 
defaults, press and release the 

s

 but-

ton. The lower display will indicate 
FCTRY.
Press and release the Power button to 
restore the factory configuration and 
restart the gauge. This does not affect 
calibration. 
For User configuration press and release 
the 

t

 button to change the lower display 

to USER. This allows you to configure 
the gauge as described in the following 
sections.
Press and release the center power but-
ton to continue with configuration.

Zero/Clear

Memory

USER

t

s

*

Zero/Clear

Memory

*

USER

- or -

t

s

*

FCTRY

User

Factory

*

or

Auto       Save      Man 

 setting

MAN

Zero/Clear

Memory

t

s

*

r

l

c

ERROR INDICATIONS

Zero Error

Attempting to zero gauge reference mod-
els with more than approx. 3% of full-
scale pressure or vacuum applied will 
result in an error zero condition. The 
display will indicate 

Err 0

If your intent was to zero the gauge, make sure to remove all 
pressure from the gauge port. 
Press the power button to reset the error.

Excessive Vacuum

Excessive vacuum applied to a pressure-
only gauge will display as 

-Err 

until the 

vacuum is released. Note: A pressure-
only gauge will be damaged by excessive 
vacuum.

Excessive Pressure

If 112.5% over range pressure is applied, 
an out-of-range indication of 

1 – – –

 or 

1.–.–.–

 will be displayed depending on 

model. Use gauge within the appropriate 
pressure range to prevent damage.

0

r

r

E

r

r

E

- or -

Summary of Contents for CTP3B

Page 1: ...CD Calibration data 5 test points and date NC NIST traceability documentation 5 points and date Vacuum Range Resolution Selectable Units CTP3B15PSIVAC 0 to 14 70 psig vacuum psi inHg torr mmHg inH2O ftH2O oz in2 cmH2O g cm2 kg cm2 atm bar mbar MPa kPa Vacuum Pressure Range Resolution Selectable Units CTP3B 15PSIG 14 70 to 15 00 psig psi inHg psi inHg torr mmHg inH2O ftH2O oz in2 cmH2O g cm2 kg cm2...

Page 2: ...e gauge 4 Due to the hardness of 316 stainless steel it is recom mended that a thread sealant be used to ensure leak free operation 4 For contaminated media use an appropriate screen or filter to keep debris out of gauge port 4 Remove system pressures before removing or installing gauge 4 Install or remove gauge using a wrench on the hex fitting only Do not attempt to turn gauge by forcing the hou...

Page 3: ...nd move to the next parameter Zero Clear Memory PSIG INHG Zero Clear Memory t s or t s Auto Shutoff Time Selection The auto shutoff time is displayed on the upper display The lower display will indicate AST M for minutes or AST H for hours Use the s and t buttons to select 0 manual shutoff 1 2 5 10 15 20 or 30 minutes or 1 2 4 or 8 hours A setting of zero disables the auto shutoff timer Use the Po...

Page 4: ...he available engi neering units as well as percent PCT For greatest accuracy use the s and t buttons to select engi neering units for calibration with highest number of display counts Press and release the Power button when the appropriate engineering units are displayed Continued Absolute Reference Gauges Apply full vacuum to the gauge The character display will alternate between ZERO and CAL Pre...

Reviews: