background image

SKU 54949         For technical questions, please call 1-800-444-3353         PAGE 5

8.

Diaphragm:  The Diaphragm (9) is placed in the light path to alter the amount of
light that reaches the condenser within the Microscope (for enhancing contrast in
the image).  (See Assy. Diagram.)

9.

Base:  The Base (12) supports the weight of all of the Microscope parts.
(See Assy. Diagram.)

Specimen Preparation:

1.

When observing a specimen by transmitted light, light must pass through the
specimen in order to form an image.  The thicker the specimen, the less light
passes through.  The less light that passes through, the darker the image.

2.

The specimens must be thin (0.1 to 0.5mm).

3.

Many specimens must be cut into thin sections before observation.  Specimens
such as rock or semiconductors are too thick to be sectioned and observed by
transmitted light, so they are observed by the light reflected from their surfaces.

Microscopic Terms:

1.

Depth of Field:  The vertical distance, from above to below the focal plane, that
yields an acceptable image.

2.

Field of View:  The area of the specimen that can be seen through the Micro-
scope with a given Objective Lens (5).

3.

Focal Length:  The distance required for an Objective Lens (5) to bring the light
to a focus (usually measured in microns).

4.

Focal Point/Focus:  The point at which the light from an Objective Lens (5)
comes together.

5.

Magnification:  The product of the magnifying powers of the Objective Lenses
(5) and Zoom Eyepiece (1).  The numbers marked on the Objective Lenses
indicates how many times the specimen on the Slide (7) is being magnified.  The
Objective Lenses on this Microscope allow a specimen to be magnified from
eight (8X) to twenty (20X) times its actual size.

Summary of Contents for 54949

Page 1: ... portion of this manual or any artwork contained herein may be reproduced in any shape or form without the express written consent of Harbor Freight Tools For technical questions please call 1 800 444 3353 TO PREVENT DAMAGE TO THE PRODUCT READ AND UNDERSTAND ALL WARNINGS AND INSTRUCTIONS BEFORE USE Due to continuing improvements actual product may differ slightly from the product described herein ...

Page 2: ...IMPORTANT SAFETY INSTRUCTIONS CAUTION READ AND UNDERSTAND ALL INSTRUCTIONS Failure to follow all instructions detailed in this manual may result in damage to the Microscope SAVE THESE INSTRUCTIONS SKU 54949 For technical questions please call 1 800 444 3353 PAGE 2 WORK AREA 1 Keep your work area clean and well lit Cluttered and dark work areas invite accidents Item Description Total Magnifications...

Page 3: ...ecise optical instrument and should be handled carefully and kept clean and dry to avoid damage When not in use keep the Microscope covered with a dry clean cloth Dust and moisture can damage the instrument 5 Do not touch the optical parts of the Microscope with your fingers Doing so may damage the lenses 6 If necessary clean the lenses with a soft dry brush Do not clean with moist cleaning agents...

Page 4: ...ounder high magnifica tion this Microscope can bring increasingly smaller areas into view See Assy Diagram 2 Drawtube Zoom Eyepiece The Drawtube 1 keeps the Eyepiece 1 at the proper distance from the Objective Lens 5 and blocks out stray light The Zoom Eyepiece transmits and magnifies the image from the Objective Lens to your eye The Zoom Eyepiece is also used to make fine adjustments to focus the...

Page 5: ...semiconductors are too thick to be sectioned and observed by transmitted light so they are observed by the light reflected from their surfaces MicroscopicTerms 1 Depth of Field The vertical distance from above to below the focal plane that yields an acceptable image 2 Field of View The area of the specimen that can be seen through the Micro scope with a given Objective Lens 5 3 Focal Length The di...

Page 6: ...men from 0 1 to 0 5mm thickness Then carefully position the specimen in the middle of the Slide 7 See Assy Diagram OPERATING INSTRUCTIONS 9 Make sure to store the Microscope in a clean dry safe location out of reach of unauthorized people 3 Thoroughly clean and dry the Slide 7 See Assy Diagram 5 Center the Slide 7 on the Specimen Stage 8 and hold the Slide in place using the Clip 6 See Assy Diagra...

Page 7: ... Diaphragm 4 Turret 10 Mirror 5 Objective Lens 11 Hand Wheel 6 Clip 12 Base PARTS LIST ASSEMBLY DIAGRAM NOTE Some parts are listed and shown for illustration purposes only and are not available individually as replacement parts DRAWTUBE ZOOM EYEPIECE 1 DRAWTUBE HOLDER 2 HINGE 3 TURRET 4 OBJECTIVE LENS 5 CLIP 6 SLIDE 7 SPECIMEN STAGE 8 DIAPHRAGM 9 MIRROR 10 HAND WHEEL 11 BASE 12 ...

Page 8: ...sequential damages arising from the use of our product Some states do not allow the exclusion or limitation of incidental or consequential damages so the above limitation of exclusion may not apply to you THIS WARRANTY IS EXPRESSLY IN LIEU OF ALL OTHER WARRANTIES EXPRESS OR IMPLIED INCLUDING THE WARRANTIES OF MERCHANTABILITY AND FITNESS To take advantage of this warranty the product or part must b...

Reviews: