4 - 1
4. Standard Operating Procedure
4-1. Operation Mode
Four operation modes are available with this Segmentation System:
- Local mode
- Remote modes;
(1) Remote switch mode
(2) Remote circuit mode
(3) Remote command mode
(a) Local Mode
Local mode is a mode that the machine operates by itself without any external device.
Default mode is the local mode. A cycle of sealing operation is initiated by pressing the
trigger lever at #1 sealing head unit. All the other trigger levers do not work at all.
(b) Remote Switch Mode
Remote switch mode is a mode that the machine is controlled by an external switch.
When a connector, which should have been specially pre-wired and an external switch
should have been wired to, is connected onto the remote control connector of the
control unit, the machine works in remote switch mode.
In this mode, sealing starts by pressing the remote switch. Before the switch pressed,
the trigger lever at #1 sealing head should have been pressed by tubing. If tubing is not
loaded in #1 sealing head, sealing does not start. The trigger lever at #1 head senses
presence of tubing, and then enables circuit to accept command given by the remote
switch only when tubing has been loaded in #1 sealing head unit. All the other trigger
levers do not work at all.
(c) Remote Circuit Mode
Remote circuit mode is a mode that the machine is controlled by an external signal.
When a connector, which should have been specially pre-wired with a circuit, is
connected onto the remote control connector of the control unit, the machine works in
remote circuit mode.
In this mode, sealing starts by giving a signal from the remote circuit. Before the signal
is given, the trigger lever at #1 sealing head should have been pressed by tubing. If
tubing is not loaded in #1 sealing head, sealing does not start. The trigger lever at #1
head senses presence of tubing, and then enables circuit to accept command given by
the remote circuit only when tubing has been loaded in #1 sealing head unit. All the
other trigger levers do not work at all.
In remote circuit mode, a confirmation signal is given to the remote circuit to confirm
that the sealing cycle has ended. If an indication lamp is prepared in the remote circuit,
the lamp will act as followings:
Summary of Contents for MS930A
Page 7: ...1 2...
Page 10: ...2 1 2 Appearance and Function 2 1 Control Unit MS950A E shown...
Page 13: ...2 4 2 2 Segmentation Unit MS950A E shown...
Page 16: ...3 2 Figure 3 1 Overall System Block Diagram...
Page 30: ...5 4...
Page 31: ...5 5 Figure 5 1 Troubleshooting Flow Chart Page 2 3...
Page 32: ...5 6 Figure 5 1 Troubleshooting Flow Chart Page 3 3...
Page 40: ...6 1 6 Schematic Diagram Figure 6 1 System wiring diagram...
Page 41: ...6 2...
Page 42: ...6 3...
Page 44: ...7 2...
Page 45: ...7 3...
Page 46: ...7 4...
Page 47: ...7 5...
Page 48: ...7 6...
Page 49: ...7 7...
Page 50: ...7 8...
Page 51: ...7 9...
Page 53: ...8 2...
Page 54: ...8 3...