E1200 HT 260-3
2BSafety
23.04.07
3-4
00401010_41087
Safety Device
Function
monitoring of process tube
ventilation
To prevent overpressure when ventilating the process
tube, a sensor system compares the pressure within and
outside the process tube and opens an over-pressure
valve if necessary. If opening the over-pressure valve
fails to prevent a limiting value from being exceeded
within the process tube, the inert gas flow is interrupted
for the corresponding process tube. Protection for
persons and property.
horizontal furnace casing
Protection against burns and electrocution. Protection for
persons.
emergency inert gas
purging
Purges the process tube in the case of power loss or
pressure loss. Protection for persons and property.
pressure monitoring cooling
water circuit
Monitors the pressure within the cooling water circuit. If
the pressure in the cooling water circuit exceeds beyond
8 bar, the overpressure valve opens. Protection for
persons and property.
flow monitoring in the outlet
line of the cooling water
circuits
Monitors the cooling water outlets. Switches the heating
currents of all heating cassettes off when errors occur.
Reports errors to the control system centrotherm
machine safety (CMS). Protection for persons and
property.
monitoring of leak water tub Monitors cooling system leaks. Reports errors to the
CMS control system. Protection for persons and property.
heating cassette casing
Protection against burns and electrocution. Protection for
persons.
heating zone temperature
monitoring
Limits the process tube temperature. Switches the
heating current of the corresponding heating cassette off
when errors occur. Reports errors to the CMS control
system. Protection for persons and property.
temperature monitoring of
the heating cassette's
outside wall
Monitors the temperature on the outside wall of the
heating cassette. Reports errors to the CMS control
system. In case of overtemperature, CMS triggers the
switch-off of the heating current of the corresponding
heating cassette. Protection for persons and property.
temperature monitoring of
power controller (solid state
relay)
Switches the heating current off in case of
overtemperature. Reports errors to the CMS control
system. The CMS triggers the switch-off of the heating
currents of the heating cassette. Protection for persons
and property.
temperature monitoring of
transformer
Monitors the temperature of the heating transformers.
Reports errors to the CMS control system. Protection for
persons and property.
monitoring of main fan
Reports errors concerning breakdown to the CMS control
system. Protection for persons and property.
movement stop/handling
stop (option)
Stops all mechanical movements. Protection for persons
and property.
Summary of Contents for E1200 HT 260-3
Page 1: ...Operating Instructions for Horizontal Furnace Equipment E1200 HT 260 3 Version 1 0 1_41087 11 ...
Page 2: ...E1200 HT 260 3 23 04 07 00401010_41087 ...
Page 4: ...E1200 HT 260 3 Document Information 23 04 07 00401010_41087 ...
Page 16: ...E1200 HT 260 3 Index 23 04 07 00401010_41087 ...
Page 18: ...E1200 HT 260 3 List of Figures 23 04 07 00401010_41087 ...
Page 20: ...E1200 HT 260 3 List of Tables 23 04 07 00401010_41087 ...
Page 26: ...E1200 HT 260 3 0BOperating Instruction Remarks 23 04 07 1 6 00401010_41087 ...
Page 34: ...E1200 HT 260 3 1BIntroduction 23 04 07 2 8 00401010_41087 ...
Page 50: ...E1200 HT 260 3 2BSafety 23 04 07 3 16 00401010_41087 ...
Page 78: ...E1200 HT 260 3 4BTransport 23 04 07 5 4 00401010_41087 ...
Page 180: ...E1200 HT 260 3 8BMaintenance 23 04 07 9 76 00401010_41087 ...
Page 228: ...E1200 HT 260 3 10BSpare and Wearing Parts 23 04 07 11 2 00401010_41087 ...