MESYS XL Manual
2.3.
Norms & Guidelines
This device has been tested and approved for compliance according
Class 1 Group B for industrial instruments. cetoni GmbH declares its
exclusive responsibility that this product is consistent with the relevant
provisions stated at the end of this document.
Operation is to conform to the following requirements:
1.
The device must not cause harmful interfering radiation.
2.
The device must be able to handle interfering radiation, including radiation
that may cause unintended operation.
The device complies with critical values according to EN55011 Class 1 Group B.
The device has passed the tests according to DIN EN 61000-4-4 (Burst) and DIN
EN 61000-4-5 (Surge).
2.4.
Application
2.4.1.
General Device Description
cetoni's
NE
MESYS devices are syringe pumps. They allow the filling and emptying
of syringes based on a relative linear movement between a syringe body fixture
and a syringe piston fixture.
2.4.2.
Intended Use
The
NE
MESYS XL ultra-high pressure module is for the precise and pulsation-
limited dosage of fluid streams with an operating pressure of up to 890 bar and
flow rates of nanoliters to mililiters per second.
Its default operating environment is a laboratory-like room.