All XL syringes feature a 24 mm hex nut directly at the screw-in thread. Use a
24 mm wrench to tighten the syringe fit by half to three quarters of a turn.
To re-mount a syringe, first hand-tight fit the syringe to the holder. Using the
wrench, an additional one quarter turn should suffice to ensure leak-
tightness of the connection.
4.
Connect the syringe piston to the piston holder using the supplied hexagon
socket. Tighten the screw using a 4 mm Allen key.
5.
Replace the safety cover by placing its two hooks into the syringe holder.
Click into place by pushing at the end near the piston head (see figure
below).
Hex nut
Syringe piston
Piston holder
Hexagon socket