CAUTION
When used over extended periods of time, the syringe may display
wear. Please check the condition of the syringe regularly and exchange
as required.
CAUTION
Please ensure that the device has been switched off using the software
prior to mounting or exchanging syringes.
5.1.3.
Disassembly & Maintenance of Syringes
The body of a
NE
MESYS XL syringe consists of the syringe cylinder plus a front and
a rear cover. The actual configuration differs slightly depending on the size
(volume) of the syringe. The following are specific descriptions for each syringe
size of how to carry out syringe maintenance:
1.
For the 10 ml syringe, both cylinder covers are directly screwed onto the
cylinder; remove them either directly or by using a 24 mm spanner wrench.
The front cover contains an O-ring seal, the rear cover an O-ring seal, a
support ring, and a support disc. After both covers have been removed,
these parts can easily be disassembled and replaced or cleaned. To
reassemble, simply carry out the same steps in reverse order.