CAUTION
Please ensure that all used pipes and tubing meet the expected pressure
ratings.
The following steps guide you through the handling of the screw connection
between the syringe socket and the fluidic conduit.
5.2.1.
Initial Swagelok
Setup
1.
Fully insert the tube / hose into the fitting and against the shoulder; rotate
the union nut finger-tight.
2.
Mark the union nut at the 6 o’clock position.
3.
Tighten the union nut three-quarters turn to the 3 o’clock position with an
open-end wrench.
5.2.2.
Disassembly
1.
Prior to disassembly, mark a line along the union nut and fitting body. Use
these marks to ensure you return the nut to the previously pulled-up
position.
2.
Loosen the union nut with an open-end wrench.
3.
Pull the tube / hose out. The union nut and the ferrules remain on the tube /
hose.
Cap nut
G1/4" thread