5.2.3.
Reassembly
1.
To reassemble, insert the tube / hose with preassembled ferrules into the
fitting body until the front ferrule seats against the fitting body.
2.
Rotate the union nut with open-end wrench to the previously pulled-up
position as indicated by the marks you made before; at this point you will
feel a significant increase in resistance.
3.
Tighten the union nut slightly.
ATTENTION
You should check all fluidic connections for tightness following each
(re-)assembly and at regular intervals during operation.
5.3.
Signal Interface
The
NE
MESYS XL module comes with a signal interface, labeled
Sensor
at the
module's front plate. It allows you to connect for example two pressure sensors
to be placed at various points within your application.
The pin assignment of the receptacle is as detailed in the following table.
A suitable plug connector is available from Hirose (order number:
HR10A-10P-12P(73)). Of course, cetoni will also provide you with suitable
peripheral devices.
Signal Interface