NE
MESYS XL Manual
35
PIN
Signal
Cable Colour
Description
1
Analog Input 1
Brown
12-bit ±10 V (differential)
2
Analog Input 2
Yellow
12-bit ±10 V (differential)
3
Digital Input 1
Orange
+9…+24 VDC (Ri=1,8kΩ)
4
Digital Input 2
Green
+9…+24 VDC (Ri=1,8kΩ)
5
Digital Input 3
Blue
+9…+24 VDC (Ri=1,8kΩ)
6
Digital Output 1
Violet
24 VDC; 20 mA; Optocoupler
7
Digital Output 2
Grey
24 VDC; 20 mA; Optocoupler
8
Digital Output 3
White
5 VDC; 60 mA; Totem-Pole
9
Digital Ground
Black
10
+24 V Out
Ret
+24 VDC / <1 A
11
+5 V Out
White
+5 VDC / <150 mA
Brown
12
Analog Ground
White
Black
The software configuration of pressure sensors is described in the software
manual. Please read and follow the description carefully before connecting and