Ready for Use
6
3. Perform testing on worktables with grounded and conductive mats; wear antistatic ankle strap
when working in areas with conductive ground surface. You’d better wear antistatic ankle strap
whether you could confirm the conductivity of ground surface or not.
2.2 Startup and Shutdown
Before powering on 3656 Series Vector Network Analyzer, please check power supplying
equipment as required in 2.1.2
“Power Supply Requirements”
. Do not power on instrument until
confirming that power supplying device is in normal condition.
2.2.1
Startup
Connect AC POWER with three-core power cord in conformity with requirements of international
security standard. Press
【
ON/STANDBY
】
key in bottom left corner on front panel. Analyzer
will take 40s to enter Windows system and start running main process of measurement. In order to
stabilize performance of internal parts and obtaining better testing results, it is recommended to
pre-heat for 30 minutes before measurement.
2.2.2
Shutdown
Press
【
ON/STANDBY
】
key on bottom left corner of front panel of the analyzer. Do not hold it
down for long or do not release it. The analyzer will exit from measurement application process
automatically, turn off Power of its own and enter into Standby status. Please switch power switch
on rear panel to Off or pull out power cord if required when the analyzer is under standby status.
The analyzer may sometimes fail to respond to operation of mouse or keyboard as
abnormalities of operating system or application process. if it occurs, you may turn it off by
holding
【
ON/STNDBY
】
key for at least 4s. However, it shall be avoided in normal
conditions, as unstable system might be led to. Although the analyzer will recover to normal
When the analyzer works on a rack, you must guarantee smooth cross-ventilation between
internal instrument and external instrument. Whenever thermal power inside rack increases
for 100W, ambient temperature (outside rack) is required to be 4
°
C lower than maximum
working temperature of the analyzer. If total thermal power exceeds 800W then compulsory
air cooling measures must be taken.