5. Menus
5.2 Window menu
71
Set the center frequency with the range of 13.5kHz~7.999995GHz, the default value is 98MHz, if the set
center frequency and the current bandwidth are not coordinated, the bandwidth will be automatically
adjusted to the best value that is compatible with the desired frequency.
[Span]
Set the frequency width with the range of 9kHz~7.999991GHz, the default value is 20MHz.
[Wave Detection Type]
Set the detection type, including dynamic refresh, minimum value, maximum value, and average four
trace types, and the default is dynamic refresh.
[Scan Count Mode]
Set the scan count mode, including unlimited and manual two modes, the default is unlimited mode.
[Scan Count]
Set the scan count which is valid when the scan count mode is set to manual. The setting range is
1-1000, and the default scan count is 1.
[Measurement Mode]
Set the measurement mode, including two modes: periodical and continuous, and the default is
periodical mode.
[Measurement Time Mode]
Set the type of measurement time, including automatic and manual modes. Auto mode is selected by
default (in the auto mode the measurement time is automatically calculated by the receiver).
[Measurement Time]
Set the measurement time, the range is 500μs-900s. The default value is set to 500μs and valid when
the measurement time mode is manual.
[Dwell Time Mode]
Set the dwell time mode, including unlimited and manual modes, the default is manual mode.
[Dwell time]
Set the dwell time with the range of 0~60s. The default setting is 500ms, and the dwell time mode is valid
when set to manual.
[No Signal Time State]
Set the no-signal time state, including on and off. The default setting is on.
[No Signal State Dwell Time]
Set the no-signal state dwell time. The range is 0~60s, and the default is set to 0s. No-signal time state is
valid when open.
[Frequency Step Mode]
Set the frequency step mode including the automatic and manual modes, and the automatic mode is
selected by default.
[Frequency Step Interval]
Set the frequency step interval which is valid when the frequency step mode is manual. The range is
1Hz~1GHz and the default value is set to 1MHz.
[Differential Mode]
Set the differential display of the spectrum. Differential mode will display when it is on, and it is in normal
display when it is off.
2) Display
The display property pages corresponding to the RF Panorama window in Frequency Scan mode are
the same as the display property pages of the RF Waterfall window in Section 5.2.2.3.