5. Menus
5.2 Window menu
73
The measurement property pages corresponding to the RF panorama window in list scan mode are
shown in Fig. 5.48.
Fig. 5.48 Measurement property pages corresponding to the RF panorama window in list scan mode
The measurement-related parameters are described below:
[Memory Table Squelch Level]
Set the state of the memory table squelch level, including the open and close states, the default is open.
All frequencies in the memory table use their own squelch level value when turned on, and the same
squelch level (squelch level value on the main screen) when turned off.
[Wave Detection Type]
Set the detection type, including dynamic refresh, minimum value, maximum value, and average four
trace types, and the default is dynamic refresh Type.
[Scan Count Mode]
Set the scan count mode, including unlimited and manual modes.The unlimited mode is set by default.
[Scan Count]
Set the scan count which is valid when the scan count mode is set to manual. The setting range is
1-1000, and the default scan count is 1.
[Measurement Mode]
Set the measurement mode, including two modes: periodical and continuous, and the default is
periodical mode.
[Measurement Time Mode]
Set the type of measurement time, including automatic and manual modes. Auto mode is selected by
default (in the auto mode the measurement time is automatically calculated by the receiver).
[Measurement Time]
Set the measurement time, the range is 500μs-900s. The default value is set to 500μs and valid when
the measurement time mode is manual.
[Dwell Time Mode]
Set the dwell time mode, including unlimited and manual modes, the default is manual mode.
[Dwell time]