Chapter III Vertical System
12
Chapter III Vertical System
This oscilloscope provides 4 analog input channels CH1
CH4, and each channel can be separately controlled.
The vertical system is mainly used to decrease the large signal and increase the small signal, so as to satisfy the
requirements of the ADC input scope.
Fig. 3.1 Vertical Operation Area of the Oscilloscope
When any channel key CH1
CH4 is pressed down, the menu of the corresponding channel will pop up at the
bottom of the screen:
Fig. 3.2 Channel Menu
3.1
Description of Keys/Buttons
Table 3.1 Description of Keys/Buttons
Keys/Buttons
Description
1
4
It is used to open the menu of the corresponding analog channel for channel setting
Scale
It is used to set the corresponding vertical scale of each analog channel
Setting scope: 50Ω: 1 mV/di
1 V/div, 1MΩ: 1 mV/div
10 V/div
Position
It is used to adjust the corresponding vertical offset of each analog channel, which has
“
Push to Zero
”
function.
Adjustment scope: 1 mV/div
100 mV/div: ±1 V
200 mV/div
1 V/div: ±10 V
2V/div
10V/div: ±100V
3.2
Probe
3.2.1 Probe port
There are 4 analog channels on the front panel of the 4456 series digital oscilloscope, which have 4 BNC input
connectors. To ensure reliable connection, ensure that your probe or cable connector is pushed in and locked.
This oscilloscope supports passive voltage probe, high voltage single end probe, high voltage differential probe,
current probe and active voltage probe. When the passive voltage probe, high voltage single end probe, high
voltage differential probe and current probe are used, the input impedance of the oscilloscope shall be set to 1 MΩ
and the probe attenuation coefficient shall be consistent with that of the selected probe. When the active voltage
probe is used, the input impedance of the oscilloscope shall be set to 50Ω and the probe attenuation coefficient
shall be consistent with that of the selected probe.
Fig. 3.3 Schematic Diagram of the Probe Port
3.2.2 Calibration of the standard passive voltage probe
Summary of Contents for 4456 Series
Page 10: ...VIII...
Page 55: ...Chapter V Trigger System 45 Fig 5 51 RF Test Signal Example...
Page 69: ...Chapter VII Mathematical Operation 59 Fig 7 10 Advanced Math Example...
Page 71: ...Chapter VIII Display System 61 Fig 8 3 Wfm Palette Menu Fig 8 4 Normal Palette...
Page 72: ...Chapter VIII Display System 62 Fig 8 5 Inverted Palette Fig 8 6 Temperature Palette...
Page 75: ...Chapter VIII Display System 65 Fig 8 12 XY Display Example...
Page 165: ...Chapter XI Protocol Analyzer Option 155 Fig 11 242 1553 Trigger and Analysis Example...