Chapter II Operation Guide
15
a) All instruments must be properly grounded to prevent static electricity.
b) The staff must wear anti-static wrist straps before contact with connectors or core wires or
any assembly.
c) The central conductor of the cable must be grounded before the cable is connected to the
instrument for tests. This can be realized according to the following steps: connect the
short-circuiting device to one end of the cable to short-circuit the central conductor of
the cable and the external conductor, wear the anti-static wrist strap, hold the casing of
the cable connector, connect the other end of the cable, and remove the short-circuiting
device.
5. Input/output port protection
The Radio Test Set is equipped with multiple RF/audio and other input/output ports. The
standard impedance of the RF port is 50Ω, the matching impedance of the audio input port is
optional, and the audio output port is limited by the current. Therefore, the test signal or
appropriate load impedance must be applied in strict accordance with the port requirements to
prevent the circuits at subsequent levels from damage.
6. Cleaning of display of front panel
If required after operation of certain time, the display panel of the instrument can be cleaned
according to the following steps.
a) Shut down the instrument and disconnect the power line.
b) Wipe display panel with clean and soft cotton cloth with detergent.
c) Dry display panel with clean and soft cotton cloth.
d) Do not connect power line until detergent dries out.
CAUTION:
There is an anti-static coating on the surface of the display screen. So
never use cleaning agent including fluoride, acid or alkaline substance
and do not directly spray the agent to the display panel, or the agent
may enter the monitor and further cause damage to the device.
Section 3 Battery Installation and Replacement
1. Battery description
4992A Radio Test Set is equipped with one high-capacity rechargeable lithium ion battery.
WARNING:
Each RF/audio input port of the Radio Test Set should meet
the maximum allowable input level requirements. The applied
signal must conform to the limit requirements; otherwise, the
instrument will be damaged.
!
Summary of Contents for 4992A
Page 1: ...I 4992A Radio Test Set User Manual China Electronics Technology Instruments Co Ltd...
Page 2: ......
Page 5: ......
Page 6: ......
Page 7: ......
Page 23: ...Article I Handling Instructions 11 Article I Handling Instructions...
Page 93: ...81 Article II Technical Specifications...
Page 132: ...Article III Maintenance Instructions 120 Article III Maintenance Instructions...