23
9.
Following the fluorescein reading,
Calibration Complete
will be displayed on the screen
(Figure 3.9). You may now proceed to your experiment setup.
Figure 3.9
10.
If calibration fails,
Calibration Failed
will be displayed on the screen (Figure 3.10).
Repeat the calibration process in such instances. If failure persists on repeat, contact
Technical Support.
Figure 3.10
It is not recommended to reuse the calibration dyes because photo-bleaching occurs at
different levels from well to well. If you would still like to reuse the dyes, pool the
solution first and aliquot again as before.
3.4 Open qPCR Dual Channel System Calibration
The Open qPCR dual channel system requires calibration with the FAM + HEX Calibrator Set.
The calibrators are provided in lyophilized format. Prior to calibrating the Open qPCR, ensure
that the lyophilized calibrators are reconstituted thoroughly (~ 45 minutes). The calibration
process, including calibrator reconstitution, will take approximately 75 minutes to complete.