50
CHAPTER 8 – Troubleshooting
8.1 Amplification Curves
Amplification Curves
Problem
Cause
Solution
1
Inconsistent data
where replicate
samples do not
display similar
results
a. Low enzyme
concentration
Increase enzyme
concentration or use a
different enzyme
b. Low target copy
number below the LOD for
the assay
Increase sample
concentration
c. Pipetting errors
Repeat assay
2
No amplification
a. Lack of target in sample
Confirm by testing a positive
control sample
b. Assay design failure
Try a different assay
c. Sample degradation
For RNA samples, denature
via formamide and then run
an agarose gel. For DNA
samples, run an agarose gel
or use a capillary
electrophoresis system. A
smeared gel reflects sample
degradation. If there is a
smear, prepare a fresh
sample