55
8.2
Melt Curves
MELT CURVES
Problem
Cause
Solution
1
Large quantity of
non-specific
product amplified
a. Excess primer
Decrease primer
concentration
b. Polymerase activity
begins before thermal
cycling starts
Use hot start polymerase
and ensure that reaction
tubes are kept chilled before
placing into the thermal
cycler
c. Primer design not
optimal for the assay
Run the sequence through
Primer-BLAST to check for
homology
d. Excessive hold times
Reduce the number of hold
times
8.3 Factory Reset
A factory reset will restore your machine to factory condition. Remember to back up your data
before performing a reset as this procedure will erase all data.
To perform the factory image reset:
1.
Turn off your Open qPCR. Disconnect any USB cables.
2.
Using a toothpick, press the reset button through the small hole in the rear of the
machine. See attached photo for exact location. Press the button and continue to hold it
down.
Figure 8.1