Summary of Contents for CO-OPM-10

Page 1: ...ifications 2 2 2 6 Graph 9 2 3 7 System Info 18 2 2 1 Power ON OFF 6 2 3 2 Offset 14 1 3 Accessories 3 2 2 7 Recall 10 2 3 8 Scan Display 19 2 2 2 Power Meter 7 2 3 3 Auto Power Off 15 1 4 Optical Sta...

Page 2: ...ast power up and test High quality graphics 1 2 Specifications Channels 10 CorWave Wavelength 1290 1291 1293 1295 1471 1491 1591 1611 1310 1550 nm Channel Spacing Custom Measurement Speed 0 5 sec all...

Page 3: ...Data cable 1 ea Charger 1 ea User s Manual 1 ea 1 4 Optical Standards Wavelength Parameter Unit Specification Number of Channels 10 Measuring Wavelengths nm 1290 1291 1293 1295 1471 1491 1591 1611 131...

Page 4: ...provided for charging Typical charge time is six hours Charging conditions LED Color Action of LED Conditions for battery charge Red ON Charging Green ON Fully charged 1 6 Meter Care Do not subject th...

Page 5: ...M Power levels of all wavelengths are displayed as a table SCAN All Wavelengths are scanned and power levels are displayed dB dBm Toggle between dB and dBm UOM Move UP in a menu page up in result scre...

Page 6: ...Contents CO OPM 10 p 6 2 2 Key Function 2 2 1 Power ON OFF To turn on and off Press the button for more than 2 seconds the logo shown appears The meter then displays the channel with the highest powe...

Page 7: ...the wavelength which has been detected with the highest level of power If you select Manual the power of the wavelength selected by the user is displayed in real time You can select the different wave...

Page 8: ...ack to Contents CO OPM 10 p 8 2 2 4 Scan Press the button the light source is scanned and the power level on each channel is displayed You can select the page using the arrow keys 2 2 5 dB dBm This bu...

Page 9: ...ific Optical Testing and Measurement Equipment back to Contents CO OPM 10 p 9 2 2 6 Graph The button is used after scanning to show power levels in graphical format Press the button again it returns t...

Page 10: ...com Wavelength Specific Optical Testing and Measurement Equipment back to Contents CO OPM 10 p 10 2 2 7 Recall The button will recall saved measurements Press to select the data you want to display Re...

Page 11: ...ipment back to Contents CO OPM 10 p 11 2 2 8 Save The button allows the user to save a measurement The following message shows and the measurement is saved using the current date and time as the filen...

Page 12: ...Contents CO OPM 10 p 12 2 3 MENU The button allows you to configure the operating environment of the CO OPM 10 The menu consists of SCAN SET Off Set Auto Power Off Time Set Data Delete Data Format Sy...

Page 13: ...choose the Scan Set All Scan Selected are displayed All Scan will measure all 10 wavelengths Selected is used to choose which wavelengths are measured If you press the Selected you can see the screen...

Page 14: ...t a test point you can set the Offset by pressing the key Press the button after selecting All Channel you can set up all channel s offset value between 25 99 to 25 99 dB and sign can be changed to nu...

Page 15: ...p 15 2 3 3 Auto Power Off The meter will automatically power down if the user does not input any key in the determined time Press the button after selecting Auto Power Off you can set the Auto power o...

Page 16: ...d Measurement Equipment back to Contents CO OPM 10 p 16 2 3 4 Time Set Press the button after selecting Time Set to set the date and current time Enter date in YY MM DD format Enter time in HH MM SS f...

Page 17: ...ata Delete Press the button after selecting Delete Data you can delete saved measurements Use and to select data set to be deleted Press the key after choosing the data you want to delete the message...

Page 18: ...saved files by formatting the memory Press the key after Data Format message appears on the screen to confirm Press the button all files are deleted from the memory CAUTION This option deletes ALL DA...

Page 19: ...the meter REGARDLESS of power level If you choose the Measured only wavelengths with a power level higher than 40 dBm are displayed 2 3 9 PM Mode The PM Mode consists of the Auto and Manual This can...

Page 20: ...nsists of the Single and Repeat This can be selected using If you choose the Repeat when selecting SCAN shows the real time changing power values There are approximately 3 measurements per second If y...

Page 21: ...rement Equipment back to Contents CO OPM 10 p 21 2 3 11 Threshold This will change the baseline of the meter 30 00 dBm is the default setting Press the button after selecting Threshold you can set the...

Page 22: ...ChallengerOptics com Wavelength Specific Optical Testing and Measurement Equipment back to Contents CO OPM 10 p 22 2 3 12 Graph Color You can change the graph color...

Page 23: ...shold The Below Threshold consists of the Display and Hide This can be selected using If you choose the Visible any power level BELOW the threshold will be displayed as it is measured If you choose th...