V-ACTOR HF HANDPIECE OPERATIONS, SET-UP, CLEANING, AND MAINTENANCE
EN
INTELECT
®
RPW 2 USER MANUAL
TREATMENT
SAFETY INFORMATION
Before using the device, the user must make sure it is functioning safely and in proper condition.
Each time the device is transported, subsequently make sure that all functional checks have been performed on the device
before you start treatment. For further information, reference page 69, FUNCTIONAL CHECKS.
Note: The maximum energy level used during treatment must not cause the patient undue pain under any circumstances.
CARRYING OUT TREATMENT
• Apply a sufficient amount of massage oil to the patient's skin in the treatment area and to the V-ACTOR HF vibration
transmitter.
• Perform V-ACTOR HF treatment as recommended in the application brochure/application recommendations.
*V-Actor is a registered trade mark of Storz medical AG
Caution!
The handpiece may not be operated while idling (without an impact surface).
• Do not trigger pulses unless the vibration transmitter is in contact with the treatment zone!
Caution!
Over extended periods, the noise of the pulses may be perceived as unpleasant!
• Offer ear protection to the patient.
• Recommendation: The user should also wear ear protection.