1 0 . e M c ta B l e S
81
EN
REVOLUTION WIRELESS™
Voltage dips, short
interruptions and
voltage variations
on power supply
lines
IEC 61000-4-11
<5 % V
T
(dips >95 % de U
T
)
for 0.5 cycle
<40 % V
T
(dips >60 % de U
T
)
for 5 cycles
<70 % V
T
(dips >30 % de U
T
)
for 25 cycles
<5 % V
T
(dips >95 % de U
T
)
for 5 seconds
<5 % V
T
(dips >95 % de U
T
)
for 0.5 cycle
<40 % V
T
(dips >60 % de U
T
)
for 5 cycles
<70 % V
T
(dips >30 % de U
T
)
for 25 cycles
<5 % V
T
(dips >95 % de U
T
)
for 5 seconds
The quality of the power supply should
be that of a typical commercial or
hospital environment. If the device user
requires continuous operation during
mains power cuts, it is recommend
that the Device is powered by a UPS or
a battery.
Magnetic field at
grid frequency
(50/60 Hz)
IEC 61000-4-8
3 A/m
Magnetic fields at the mains frequency
should be at a level characteristic of a
typical location in a typical commercial
or hospital environment.
NOTE :V
T
is the AC supply voltage before application of the test level.