49
Crest Factor (CF) Calculation
4.2.8
Measurement Range
1.00 to 3.50
3.51 to 5.99
6.00 to 10.00
Specified
Measurement Range
(from 5V or 5A)
0 to 100% of the measurement range
Accuracy (zero
corrected in A
DC
)
± (2% R +2cts) ± (5% R +2cts) ± (10% R +2cts)
Resolution
1ct
NOTE:
Peak values limited to 1500V or 3000A
RIPPLE in DC Calculation
4.2.9
Measurement Range
0.1 to 99.9%
100.0 to 1000%
Specified Measurement
Range (from 3A
DC
and 2V
DC
)
2 to 100% of the
measurement range
0 to 100% of the
measurement range
Accuracy
± (5% R +10cts)
Resolution
0.1
NOTE:
If one of the terms for the calculation of the RIPPLE is displayed as "OL",
or forced to zero, the RIPPLE displayed is an indeterminate value, "----".
Continuity Measurement
4.2.10
Measurement Range
0.0 to 999.9
Ω
Open-circuit Voltage
≤
3.6V
Measurement Current
550µA
Accuracy
± (1% R +5cts)
Buzzer Triggering Threshold
40
Ω
Resistance Measurement
4.2.11
Measurement Range (1)
0.0 to
999.9
Ω
1000 to
9999
Ω
10.00 to
99.99k
Ω
Specified Measurement
Range
1 to 100% of the
measurement range
0 to 100% of the
measurement range
Accuracy
± (1% R +5cts)
Resolution
0.1
Ω
1
Ω
10
Ω
Open-circuit Voltage
≤
3.6V
Measurement Current
550µA
100µA
10µA
Note (1)
Above the maximum display value, the display unit indicates "OL".
The "-" and "+" signs are not managed.
Specific Specifications in MAX/MIN mode
:
Accuracy: add 1% R to the values of the previous table
Capture of the extreme: approximately 100ms