3.1.5. USE OF ANOTHER PH ELECTRODE
The electrode supplied with the instrument has an integrated temperature probe. If you use another electrode, one without an
integrated temperature probe, you must measure the temperature of the solution.
The instrument indicates that the temperature can be modified by displaying
MTC
next to the temperature.
MTC = Manual Temperature Compensation.
You must then correct the temperature displayed, using the ▲ and ▼ keys, so that it is equal to the measured temperature of the
solution.
The instrument corrects the response of the electrode for temperature.
To calibrate the electrode, always correct the temperature first. And make sure that all of the buffer solutions are at the same
temperature.
To connect your electrode, use one of the optional connection adapters (see § 1.2). These adapters also let you connect a PT1000
probe.