14
(1) Bandwidth: 40 Hz to 5 kHz
(3) In % of the end of scale
(2) Without the leads. Resistance of the pair of leads supplied: approx. 70 m
Ω
■
When the order of magnitude is not known, place the switch on the
highest range then progressively lower to the appropriate range.
■
To get the current in
µ
A, mA or A, multiply the value read on the
appropriate scale by the reading coefficient shown in the table.
5 - DC AND AC CURRENTS
Always switch off the circuit to test before connecting the
multimeter to the circuit. If the “Fus.” light comes on, change the
faulty fuse(s) (Reminder: minimum voltage of 100 V).
Connect the leads to the multimeter and connect in series to the
circuit with:
- the red lead in the “
µ
mA” terminal, up to 1.5 mA
- the red lead in the “15 A” terminal, from 1.5 A to 15 A
Switch off the power supply to the circuit before changing range.
(1) Common to the 100 mV DC range
(2) Without the leads. Resistance of the pair of leads supplied: approx. 70 m
Ω
(3) In % of the end of scale
Limitation 10 min On, 5 min Off up to +40
°
C max
Limitation 10 min On, 5 min Off up to +40
°
C max
Do not use the A AC input on unprotected current transformers.
A D C
5 0 µ A (1) 1 5 0 µ A
1 .5 m A
1 5 m A
1 5 0 m A
1 .5 A
1 5 A
S c a le
1 0 0
1 5 b la c k
R e a d in g
c o e ffic ie n t
x 0 .5
x 1 0
x 0 .1
x 1
x 1 0
x 0 .1
x 1
V o lta g e d ro p
a t th e
te r m in a ls (2)
< 0 .1 V
< 0 .2 5 V
< 0 .3 V
< 0 .8 V
A c c u r a c y (3)
2 .5 %
5 %
P ro te c tio n
F u s e 1 .6 A H B C
Fuse 16A
H BC
A A C
(1)
1 .5 m A
1 5 m A
1 5 0 m A
1 .5 A
1 5 A
S c a le
1 5 re d
R e a d in g
c o e ffic ie n t
x 0 .1
x 1
x 1 0
x 0 .1
x 1
V o lta g e d ro p
a t th e
te r m in a ls (2)
< 0 .8 V
< 0 .9 V
< 0 .8 V
A c c u r a c y (3)
2 .5 %
5 %
P ro te c tio n
F u s e 1 .6 A H B C
Fuse 16 A
H B C