15
RESISTANCES
If the «Voltest» light comes on, unplug the leads immediately. Present voltage
is upper to 15 V
or
.
To get the resistance in Ω, multiply the reading on the Ω scale (green) by coefficient
of the selected range: x1 or x100. The zero adjusting is automatic. To keep the
battery autonomy, avoid to untouch the rotary switch on the ranges.
Ω
x1
(1)
x100
Measurement extent
5
Ω
to 10 k
Ω
500
Ω
to 1 M
Ω
Internal resistance
140
Ω
14 k
Ω
End of scale current
10 mA
150 µA
Open circuit voltage
-1.5 V
Accuracy
(3)
10 %
Permitted overload
420 V
1) Common to the continuity sound test
.
(2) In % at mid-scale
CONTINUITY SOUND TEST
Connection and specifications: idem Resistances
Place the selector switch on the
x1 function
Continuous audible beep emitted for a resistance R < 80 Ω
BATTERY CHECKING
Put the rotary switch on the function
x1 green.
Short-circuit the leads, the needle must be positioned on the 0 (green square) of
Ω green scale. In the contrary, change the battery.
4. INFLUENCES
Additional uncertainty at the § 2 and § 3 uncertainty given.
TEMPERATURE
1 %/10°C for all the ranges except 2 %/10°C for 0.1 V DC and 2.5 %/10°C for AC
current ranges.
10 %/10°C in Ω.
FREQUENCY
Voltage ranges
- In the bandwidth 45 Hz to 500 Hz : 0 %, except 1000 V range: 0.5 %
- Outside the bandwidth 45 to 500 Hz and in the limits of § 3 bandwidth: 1.5 %,
except 10 V range: 4 %
Current ranges
- In the limit of § 3 bandwidth: 1.5 %