CAPACITANCES
■
Never measure a capacitance on a live circuit.
■
Always respect the polarity of electrolytic condensors.
This type of condenser may be sensitive to temperature,
so avoid touching it during measurement.
■
Connect the leads to the multimeter and to the terminals of the condenser.
■
Place the selector switch on the capacitance meter function .
■
Automatic range selection : read the measured value
NB : The bargraph is not operational during capacitance measurement.
■
Memorise the value, if necessary, by pressing the HOLD button.
If necessary, press the following buttons :
■
RANGE to select the ranges manually
■
to light the display
■
MIN.MAX to read the min or max values
CAPACITANCES
(1) Min measurement : 400pF.
4 nF
(1)
40 nF
400 nF
4 µF
40 µF
Resolution
1 pF
10 pF
100 pF
1 nF
10 nF
Precision
Protection
500 V rms and 750 V peack
± 3 % R
± 3 % R
± 5 % reading
± 15 % R
± 400
± 25
± 10 counts
± 10
counts
counts
counts
■
Connect the leads to the multimeter and in parallel to the circuit to be tested.
■
Place the selector switch on the Hz voltage frequency function
■
Automatic range selection : read the measured value
NB : The bargraph is non-operational on frequency measurement
■
If necessary, press the following buttons
- MIN.MAX to read the max or min value
-
to light the display
FREQUENCIES