24
A AC
40 mA
400 mA
Voltage (2) drop
600 mV
4,5 V
600 mV
Digital
resolution
10 µA
100 µA
10 mA
Bargraph
resolution
1 mA
10 mA
1 A
Accuracy (4)
± 1.5 % reading ± 2 counts
± 2 % R
± 2 counts
Protection (3)
Fuse 0.4 A HBC
Fuse 12 A HBC
AC CURRENTS
10 A
(1)
A DC
40 mA
400 mA
10 A (1)
Voltage (2) drop
600 mV
4,5 V
600 mV
Digital
resolution
10 µA
100 µA
10 mA
Bargraph
resolution
1 mA
10 mA
1 A
Accuracy
± 1.5 % reading ± 1 count
± 1.5 % R
± 2 counts
Protection (3)
Fuse 0.4 A HBC
Fuse 12 A HBC
DC CURRENTS
(1) 4000 count display limited to 10A Between 5A and 10A, to avoid
overheating, limit the operating time to 10 minutes
(2) Voltage drop at the terminals for 40mA, 400mA, 10A
(3) Permitted overloads :
600mA for 30 seconds for the 400mA range
15A for 30 seconds for the 10A range
(4) Use on frequency : 40Hz to 500Hz