4.2.3 Frequency
Special reference conditions
:
150 mV < U < 600 V
0,15 A < I < 10 A
When the switch is on the Hz or Volts position, the 300 Hz filter is not operational.
When the switch is in the V
LowZ
position, the 300 Hz filter is activated for Volts and frequency
Range
600 Hz
6 kHz
60 kHz
Specified measurement range
10 - 600.0 Hz
0.01 - 6.00 kHz
0.01 - 10 kHz
Uncertainty (
±
)
0.1 % R + 2 D
0.1 % R + 2 D
0.1 % R + 2 D
Resolution
0.1 Hz
1 Hz
10 Hz
- Below 10 Hz, the value is forced to zero.
- If the detection level is insufficient, or if the current or voltage value is forced to zero, the freq. is not determined "----- ".
4.2.4 Resistance
Special reference conditions
: the input (+, COM) must not have been overloaded following the accidental application of
voltage on the input terminals when the switch is on the
Ω
or T ° position. If this is the case, the return to normal can take
about ten minutes.
Range
Specified
measurement range
Resolution
Uncertainty (
±
)
Measurement
current
Voltage in open
circuit
600
Ω
0 – 600,0
Ω
*
0,1
Ω
1 % R + 3 D
≈
1 mA
6 k
Ω
0 – 6,000 k
Ω
0,001
k
Ω
≈
120
μ
A
60 k
Ω
0 – 60,00 k
Ω
0,01
k
Ω
≈
12
μ
A
600 k
Ω
0 – 600,0 k
Ω
0,1
k
Ω
1 % R + 2 D
≈
1,2
μ
A
6 M
Ω
0 – 6,000 M
Ω
0,001
M
Ω
1,5 % R + 3 D
≈
120 nA
60 M
Ω
0 – 60,00 M
Ω
0,01
M
Ω
3 % R + 5 D
≈
30 nA
< 5 V
4.2.5 Audible continuity
Response time < 100 ms
Range
Resolution
Uncertainty (
±
)
Voltage in open circuit
Meas. current
600
Ω
0.1
Ω
Beep triggered
< 30
Ω
+ 3
Ω
< 5 V
≈
1 mA
4.2.6 Diode test
Range
Resolution
Uncertainty (
±
)
Voltage in open circuit
Meas.
current
6 V
1 mV
Beep triggered
< 40 mV + 10 mV
< 5 V
< 1,1 mA