4.2.3 AC+DC voltage
60 mV range
: The measurement of high currents or over a long period can cause certain components to heat. In this case,
a certain waiting time is needed to return to the specified metrological conditions. It is possible to check whether the offset
has returned to an acceptable value by short-circuiting the + and COM terminals. A reading (R) of < 5 D should be obtained.
Range
Specified
measurement
range
3)
Resolution
Uncertainty
DC (
±
)
Uncertainty
AC (
±
)
Bandwidth
Input
Impedance
Crest factor
40 to 400 Hz
0.4 to 10 kHz
60 mV
1)
6 - 60 mV
0.01 mV
0.8 % R
+ 10 D
1.5 % R + 15 D
≈
400 Hz
10 M
Ω
//
< 50 pF
3 at 50 mV
600 mV
60 - 600 mV
0.1 mV
0.8 % R
+ 5 D
0.8% R + 0.5%
x [F(kHz) -0.4]
R
+ 5 D
40 Hz to
10 kHz
3 at 500 mV
6 V
0.6 - 6 V
0.001 V
0.8 % R
+ 3 D
0.8% R + 0.5%
x [F(kHz) -0.4]
R
+ 3 D
3 at 5 V
60 V
6 - 60 V
0.01 V
3 at 50 V
600 V
60 - 600 V
0.1 V
3 at 500 V
1000 V
2)
60 - 1000 V
1 V
1.42 at
1000 V
1)
This range can only be accessed using the
key. Input impedance
≈
10.6 M
Ω
// 50 pF
2)
The display indicates "+OL" in excess of +1050 V and "-OL" en excess of -1050 V or 1050Vrms.
3)
From 1 kHz, the measurement must exceed 15 % of the range
- Secondary measurements and displays: frequency (AC coupling): Fmax
≤
10 kHz, min max
4.2.4 Frequency
-
Special reference conditions 150 mV < U < 600 V
When the switch is on the Hz or Volts position, the 300 Hz filter is not operational.
When the switch is in the V
LowZ
position, the 300 Hz filter is activated for Volts and frequency
Display range
600 Hz
6 kHz
60 kHz
Specified measurement range
10 - 600.0 Hz
0.01 - 6.000 kHz
0.01 - 10 kHz
Uncertainty (
±
)
0.1 % R + 1 D
0.1 % R + 2 D
0.1 % R + 2 D
Resolution
0.1 Hz
1 Hz
10 Hz
- Below 10 Hz, the value is forced to zero.
- If the detection level is insufficient, or if the current or voltage value is forced to zero, the freq. is not determined "----- ".