47
3.2 INSTRUMENT CONFIGURATION (
SET-UP)
→
Set the rotary switch to the SET-UP position.
The parameter or value configured is validated upon return to the "PUSH btn" screen.
Warning: if the switch is turned before the return to the “PUSH btn” screen, the modified data
are lost.
The table below indicates the various parameters that can be configured and their programming
sequence.
Remark:
generally, changes from "ON" to "OFF" and/or changes of the values of the parameters are
effected using the
key.
Parameter
Presses
Values
Default values
Time / Date
successive
Euro (DD/MM)
US (MM/DD)
YYYY
HH:mm
User adjustable
Type of power supply
bAtt
niMH
bAtt
Activate/deactivate automatic
shutdown
on
OFF
on
Automatic shutdown
time
01 to 59 mn
5 mn
Activate/deactivate the buzzer
on
OFF
on
Display
the
internal
parameters of the instrument
successive
serial no.
software version
date of calibration
LCD screen
Number of measurements in
"SMOOTH" mode
2 to 5
3
Printing of configuration
Printer configuration
(data rate)
300 to 9600 bauds
9600
Default configuration
see § 3.2.1
Erase memory (totally or
partially)
see § 3.6
Type of compensation of
cables (see § 3.3)
User
Std
none
Std
Reference voltage for the
calculation of I
k
see § 4.3.2
voltage measured
Value of the low current I
TEST
in
"non-tripping" measurement
6, 9 or 12 mA
see § 4.2.2
12 mA
Threshold voltage U
L
25 or 50 V
50 V
MEM
+
2
nd
+
2
nd
2
nd
+ TEST
2
nd
MEM
+ 2 x
2
nd
+ 3 x
2
nd
+
TEST
TEST
+
2
nd
MORE
MORE
x 2
MORE
x 3
MORE
x 4