4 CHARACTERISTICS
4.1
REFERENCE CONDITIONS
Quantities of influence
Reference conditions
Temperature
23°C ±2°C
Relative humidity
45% to 75%
Supply voltage
9.0V ±0.5V
Frequency range of the applied signal
45–65Hz
Sine wave
pure
Peak factor of the applied alternating signal
√2
Position of the conductor in the clamp
centred
Adjacent conductors
none
Alternating magnetic field
none
Electric field
none
4.2
CHARACTERISTICS UNDER THE REFERENCE CONDITIONS
The uncertainties are expressed in ± (x% of the reading (R) + y points (pt)).
4.2.1
DC voltage measurement
Measurement range
0.00 V to
59.99 V
60.0 V to
599.9 V
600 V to 1000V
(1)
Specified
measurement range
0 to 100% of the measurement range
Uncertainties
from 0.00 V to 5.99 V
±(1% R + 10 pt)
from 6.00 V to 59.99
V
±(1% R +3 pt)
±(1% R +3 pt)
Resolution
0.01 V
0.1 V
1 V
Input impedance
10 M
Ω
Note (1)
Above 1000V, a repetitive beep indicates that the voltage being
measured is greater than the safety voltage for which the device is guaranteed.