31
Note (1) Above the maximum display value, the display unit indicates "OL".
- The "-" and "+" signs are not managed.
Specific characteristics in MAX/MIN mode
:
•
Uncertainties: add 1% R to the values of the table above.
•
Capture time of the extrema: approximately 100ms.
4.2.7
Diode test
Measurement range
0.000 V to 3.199 V DC
Specified measurement range
1 to 100% of the measurement range
Uncertainties
± (1% R + 10 pt)
Resolution
0.001V
Measurement current
0,55 mA
Indication: junction reversed or
open-circuit
Display of "OL" when the measured
voltage >3.199V
Note : The "-" sign is disabled for the diode test function.
4.2.8
Frequency measurements
4.2.8.1
C haracteris tics in voltage
Measurement range (1)
5.0 Hz to
999.9 Hz
1000 Hz to
9999 Hz
10,00k Hz to
19,99k Hz
Specified measurement
range
1 to 100% of the
measurement
range
0 to 100% of the measurement
range
Uncertainties
± (0.4% R + 1 pt)
Resolution
0.1 Hz
1 Hz
10 Hz
4.2.8.2
C haracteris tics in current
Measurement range (1)
5.0 Hz to 999,9 Hz
1000 Hz to 1999 Hz
Specified
measurement range
1 to 100% of the
measurement range
0 to 100% of the
measurement range
Uncertainties
± (0.4% R + 1 pt)
Resolution
0.1Hz
1Hz
Note (1) - If the level of the signal is too low (U<3V or I<3A) or if the frequency is
less than 5Hz, the device cannot determine the frequency and displays
dashes "----"