32
2.
Toggle between AC or DC voltage measurement by pressing the yellow
button (in VlowZ : only AC measurement available).
3.
Connect the red test lead to the
+
terminal and the black test lead to the
COM
terminal.
4.
Measure the voltage by touching the probes to the desired test points of
the circuit:
To avoid to measure ghost voltage, choose
, impedance of the
DMM is lower in Low Z (500
k
Ω
).
There are two ranges in the mV setting. As default, the 600 mV range is selected.
Pressing RANGE switches the instrument to 60 mV to measure weak voltages.
Measuring AC, DC or AC+DC current
3.5.
1.
Turn the switch to
,
, or
.
2.
Toggle between AC or DC current measurement by pressing the yellow
key.
3.
Connect the red test lead to either A, or mA/µA terminal and black test
lead to the
COM
terminal.
4.
Break the circuit path to be measured. Then connect the test leads across
the break and apply power.
5.
Read the measured current on the display.
Do not exceed the current limits: 10A in the 10A range and 630mA in the µA and mA
range.
MTX 204, 10A range
MTX 204, µA range
Summary of Contents for Metrix MTX204
Page 2: ...FR 1 GB 22 DE 42 IT 63 ES 83 CZ 104 NL 125 PL 146 RO 167 RU 188 ANNEX 210...
Page 7: ...5 Description 2 4...
Page 28: ...26 Device description 2 4...
Page 48: ...46 Beschreibung 2 4...
Page 69: ...67 Descrizione 2 4...
Page 89: ...87 Descripci n 2 4...
Page 110: ...108 bezkontaktn detekce nap t Popis p stroje 2 4...
Page 131: ...129 Beschrijving 2 4...
Page 152: ...150 Opis 2 4...
Page 173: ...171 Descriere 2 4...
Page 191: ...189 1 MTX 204 1 1 600 IV 1000 III 61010 2 033 61010 031 61010 2 032 1 1 1 DEEE 2002 96 CE...
Page 192: ...190 FUSED 1 1 2 IV III II...
Page 193: ...191 2 2 1 2 AA 1 5 1 5 1 5 2 2 III IV www chauvin arnoux com 2 3 MTX 204...
Page 194: ...192 2 4...
Page 195: ...193 1 2 5 2 Range 3 4 HOLD NCV 5 REL 6 Hz 7 Max Min 8 Commutateur 9 10 10 11 12 600...
Page 196: ...194 2 5 1 8 A 2 9 3 10 4 11 5 12 6 13 7...
Page 197: ...195 2 6 1 2 Multifix P01102100Z 3 4 2 3 4...
Page 198: ...196 3 3 1 61010 31...
Page 199: ...197 3 2 1 2 2 1 5 3 OL V 3 3 2 2 NCV...
Page 200: ...198 AUTO Range 3 4 1 10 500 2 VlowZ 3 COM 4 500 600 Range 60...
Page 201: ...199 3 5 1 2 3 A mA A COM 4 5 10A 10A 630 MTX 204 10 A MTX 204 3 6...
Page 202: ...200 1 2 COM 3 50 600 OL 3 7 600 OL 3 8...
Page 203: ...201 1 2 3 4 5 3 OL 3 9 1 2 COM 3...
Page 204: ...202 4 5 100 15 NCV 3 10 NCV www chauvin arnoux com 230 50 10 NCV 1 2 3 EF 4 220 230 50...
Page 205: ...203 NCV NCV 3 11 VlowZ 500 3 12 REL VLowZ Hz Hz 1 Max Min Max Min Max Min...
Page 206: ...204 4 23 C 3 C 45 75 1 5 AA Ni MH 45 65 40 1...
Page 210: ...208 6 6 1 OFF 6 2 10A OFF 1 2 3 A mA A 4 A 000 0 000 2...
Page 211: ...209 5 OL 6 6 3 LR6 F1 630 1000 10 6 3 32 F2 10 1000 30 10 38 7...
Page 214: ...212 X04836A00 Ed 02 10 2018...