65
6.2.4.3.
Intrinsic uncertainty
The intrinsic uncertainties of the current and phase measured by the sensor must be added to the intrinsic uncertainties of
the instrument for the quantity concerned (power, energy, power factor, tan
Φ
, etc.).
The following specifications are considered to be in the conditions of references of the current sensor.
Current sensors with 1 V output at Inom specifications
Sensor type
I nominal
Current
(RMS or DC)
Intrinsic
uncertainty
at 50/60 Hz
Intrinsic
uncertainty
on
ϕ
at 50/60 Hz
Typical
uncertainty on
ϕ
at
50/60 Hz
Typical
uncertainty
on
ϕ
at 400 Hz
PAC93
clamp
1000 A
dc
[1 A; 50 A[
± 1.5% ± 1 A
-
-
[50 A; 100 A[
± 1.5% ± 1 A
± 2.5°
-0.9°
- 4.5°@ 100 A
[100 A; 800 A[
± 2.5%
± 2°
- 0.8°
[800 A; 1000 A[
± 4%
- 0.65°
C193
clamp
1000 A
ac
[1 A; 50 A[
± 1%
-
-
[50 A; 100 A[
± 0.5%
± 1°
+ 0.25°
+ 0.1°@ 1000 A
[100 A; 1200 A[
± 0.3%
± 0.7°
+ 0.2°
MN93
clamp
200 A
ac
[0,5 A; 5 A[
± 3% ± 1 A
-
-
-
[5 A; 40 A[
± 2.5% ± 1 A
± 5°
+ 2°
- 1.5°@ 40 A
[40 A; 100 A[
± 2% ± 1 A
± 3°
+ 1.2°
- 0.8°@ 100 A
[100 A; 240 A[
± 1% + 1 A
± 2.5°
± 0.8°
- 1°@ 200 A
MN93A
clamp
100 A
ac
[200 mA; 5 A[
± 1% ± 2 mA
± 4°
-
-
[5 A; 120 A[
± 1%
± 2.5°
+ 0.75°
- 0.5°@100 A
5 A
ac
[5 mA; 250 mA[
± 1.5% ± 0,1 mA
-
-
-
[255 mA; 6 A[
± 1%
± 5°
+ 1.7°
- 0.5°@ 5 A
E3N
clamp
100 A
ac/dc
[5 A; 40 A[
± 4% ± 50 mA
± 1°
-
-
[40 A; 100 A[
± 15%
± 1°
-
-
10 A
ac/dc
[50 mA; 10 A[
± 3% ± 50 mA
± 1.5°
-
-
J93
clamp
3500 A
ac
5000 A
dc
[50 A; 100 A[
± 2% ± 2.5 A
± 4°
-
-
[100 A; 500 A[
± 1.5% ± 2.5 A
± 2°
-
-
[500 A; 3500 A[
± 1%
± 1.5°
-
-
]3500 A
dc
; 5000 A
dc
[
± 1%
-
-
-
5A / Essailec
®
Adapter
5 A
ac
[5 mA; 250 mA[
± 0.5% ± 2 mA
± 0.5°
-
-
[250 mA; 6 A[
± 0.5% ± 1 mA
± 0.5°
Table 23
Summary of Contents for pel 102
Page 1: ...Power energy logger EN User s manual PEL 102 PEL 103 PEL 104 ...
Page 24: ...24 Aggregation period Date and time IP address scrolling Wi Fi address scrolling PEL104 ...
Page 36: ...36 Three phase 3 wire balanced 3P 3W b U12 U23 U31 f P Q S PF P Q S tan ϕ I1 I2 I3 ϕ I1 U12 ...
Page 38: ...38 Three phase 4 wire wye balanced 3P 4WYb I1 I2 I3 V1 V2 V3 VN U12 U23 U31 f P Q S tan ϕ ...
Page 39: ...39 DC 2 wire dC 2W DC 3 wire dC 3W P Q S PF ϕ I1 V1 P Q S tan ϕ P I V VN I1 I2 IN ...
Page 40: ...40 DC 4 wire dC 4W P I1 I2 I3 IN V1 V2 V3 VN V1 V2 VN ...
Page 48: ...48 Two phase 3 wire 1P 3W I1 I2 V1 V2 U12 VN P Q S P Q S ...
Page 51: ...51 P Q S ...
Page 85: ...85 ...