69
6.8. ELECTROMAGNETIC COMPATIBILITY
Emissions and immunity in an industrial setting compliant with IEC/EN 61326-1 or BS EN 61326-1.
With the AmpFlex
®
and the MiniFlex
®
, the typical influence on the measurement is 0.5% of full scale, with a maximum of 5A.
6.9. RADIO EMISSION
The devices are compliant with the 2014/53/EU RED directive and FCC Regulations.
https://www.chauvin-arnoux.com/COM/CA/doc/Declaration_of_conformity_PEL102.pdf
https://www.chauvin-arnoux.com/COM/CA/doc/Declaration_of_conformity_PEL103.pdf
https://www.chauvin-arnoux.com/COM/CA/doc/Declaration_of_conformity_PEL104.pdf
FCC certification
Bluetooth
FCC QOQWT11u
Wi-Fi (PEL104)
FCC QOQWF121
3G (PEL104)
FCC XPY-LISAU200
6.10. MEMORY CARD
The PEL accepts FAT32-formatted SD, SDHC and SDXC cards up to a capacity of 32GB.
The SDXC cards must be formatted in the instrument.
Number of insertions and withdrawals: 1,000.
The transfer of a large quantity of data may take a long time. Moreover, some computers may have difficulty processing such large
quantities of information, and spread sheets accept only a limited quantity of data.
We recommend optimizing the data on the SD card and recording only the necessary measurements. For guidance, a 5-day
record, with an aggregation time of 15 minutes, a record of the “1s” data and the harmonics on a three-phase four-wire network
occupies approximately 530MB. If the harmonics are not essential and if recording of them is deactivated, the size is reduced to
approximately 67MB.
The maximum durations of records for a 2GB card are the following:
7 days for recording with an aggregation time of 1 minute, the “1s” data, and the harmonics;
1 month for recording with an aggregation time of 1 minute, the “1s” data, but no harmonics;
1 year for recording with an aggregation time of 1 minute.
Do not exceed 32 records on the SD card.
For records that are long (duration greater than one week) or include the harmonics, use class 4 or higher SDHC cards.
Do not use the Bluetooth, the
Wi-fi or the 3G-UMTS/GPRS
link to upload large records: it would take too long. If no other link is
possible
, shrink the record by removing the “1s” data and the harmonics. Without these last, a 30-day record occupies only 2.5MB.
On the other hand, uploading by USB or Ethernet link can be acceptable, depending on the length of the record and the transmission
rate. To transfer the data more rapidly, use the SD card/USB adapter.
Summary of Contents for pel 102
Page 1: ...Power energy logger EN User s manual PEL 102 PEL 103 PEL 104 ...
Page 24: ...24 Aggregation period Date and time IP address scrolling Wi Fi address scrolling PEL104 ...
Page 36: ...36 Three phase 3 wire balanced 3P 3W b U12 U23 U31 f P Q S PF P Q S tan ϕ I1 I2 I3 ϕ I1 U12 ...
Page 38: ...38 Three phase 4 wire wye balanced 3P 4WYb I1 I2 I3 V1 V2 V3 VN U12 U23 U31 f P Q S tan ϕ ...
Page 39: ...39 DC 2 wire dC 2W DC 3 wire dC 3W P Q S PF ϕ I1 V1 P Q S tan ϕ P I V VN I1 I2 IN ...
Page 40: ...40 DC 4 wire dC 4W P I1 I2 I3 IN V1 V2 V3 VN V1 V2 VN ...
Page 48: ...48 Two phase 3 wire 1P 3W I1 I2 V1 V2 U12 VN P Q S P Q S ...
Page 51: ...51 P Q S ...
Page 85: ...85 ...